Prospective, randomized comparison of electromyographic neuromuscular responses from the adductor pollicis and abductor digiti minimi muscles during rocuronium-induced neuromuscular block.
György Nagy, László Asztalos, Réka Nemes, Szabolcs Lengyel, Adrienn Pongrácz, Orsolya Farkas, Sorin J Brull, Béla Fülesdi
Department of Anesthesiology and Intensive Care, University of Debrecen, Faculty of Medicine, Debrecen, Hungary. · HUN-REN Centre for Ecological Research, Institute of Aquatic Ecology, Debrecen, Hungary.
Key points
From the paper's conclusion- The present study underscores the equivalency between ADM and AP muscles for routine neuromuscular monitoring during surgery. When monitoring the non-dominant arm, the ADM and AP muscles have similar recoveries. In the dominant arm, the ADM muscle recovery is significantly faster than that of the AP muscle.
Abstract
BACKGROUND: Many clinicians monitor the abductor digiti minimi (ADM) muscle during surgery using electromyography (EMG) for ease and consistency of electrode placement, while practice guidelines recommend monitoring of responses at the adductor pollicis (AP) muscle. We compared the onset time of rocuronium neuromuscular block and the sugammadex-induced recovery time recorded simultaneously from the AP and ADM muscles. METHODS: In 44 consenting patients undergoing general anesthesia, simultaneous EMG recording of ADM and AP muscles was performed. Right- and left-hand selection of monitored muscle (AP vs. ADM) was randomized, and patients' handedness was recorded. The primary outcomes were the onset time of neuromuscular block (administration of rocuronium until train-of-four (TOF) count =0); recovery time (time from sugammadex administration until TOF ratio ≥0.9); and time to maximal recovery (sugammadex administration until TOF ratio=1.0). RESULTS: There were no significant differences between ADM and AP measurements in onset time, recovery to TOF ratio ≥0.9, or time to maximal recovery (TOF ratio=1.0) between EMG responses recorded simultaneously from ADM and AP muscles. However, a difference in recovery times between ADM and AP existed, depending on which hand was monitored. When ADM was monitored on the right hand, the readings showed earlier recovery than AP, and the mean difference was 93 s (25% faster) and 105 s (20% faster) for time to TOF ratio ≥0.9 and time to TOF ratio=1.0, respectively. We found no difference between the AP and ADM muscles when monitoring was performed on the left hand. CONCLUSIONS: The present study underscores the equivalency between ADM and AP muscles for routine neuromuscular monitoring during surgery. When monitoring the non-dominant arm, the ADM and AP muscles have similar recoveries. In the dominant arm, the ADM muscle recovery is significantly faster than that of the AP muscle.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Accuracy and Precision of Acceleromyography and Electromyography at the Adductor Pollicis Muscle in Restricted Thumb Movement Positions: An Observational Study.Anaesthesia, critical care & pain medicine · 2026
- Usefulness of intraoperative neuromonitoring by evaluation of motor evoked potentials from the abductor digiti minimi muscle elicited by transcranial electrical stimulation.World neurosurgery · 2026
- Neuromuscular Monitoring of Rocuronium 0.6 mg/kg in Children Aged 3–24 Months With Electromyography and Acceleromyography: An Observational StudyPediatric Anesthesia · 2026
- Comparison of kinemyography and electromyography during onset of and recovery from non-depolarising neuromuscular blockade: prospective observational study.Anaesthesiology intensive therapy · 2026
- Comparison of rocuronium‑sugammadex and mivacurium on recovery efficiency after pediatric isolated adenoidectomy: a propensity score‑matched cohort studyBMC Anesthesiology · 2026
- STRIDE: A Flexible, Depth-Controlled Intramuscular EMG Platform for Small-Animal Studies.IEEE transactions on neural systems and rehabilitation engineering : a publication of the IEEE Engineering in Medicine and Biology Society · 2026