Anaesthesia, critical care & pain medicine

Accuracy and Precision of Acceleromyography and Electromyography at the Adductor Pollicis Muscle in Restricted Thumb Movement Positions: An Observational Study.

Maria A Estevez, Mariel Maramba, Yaohua Ma, Sorin J Brull, J Ross Renew

Department of Anesthesiology and Perioperative Medicine, Mayo Clinic, Jacksonville, Florida, United States of America. Electronic address: estevez.maria@mayo.edu. · Department of Anesthesiology and Perioperative Medicine, Mayo Clinic, Jacksonville, Florida, United States of America.

Observational

Key points

From the paper's conclusion
  • EMG provided more accurate and precise TOFR measurements than AMG, including normalized AMG, with fewer failures, supporting EMG-based monitoring when upper extremity mobility is restricted.

Abstract

BACKGROUND: Quantitative neuromuscular monitoring commonly uses acceleromyography (AMG) or electromyography (EMG). Acceleromyography requires unrestricted thumb movement, whereas EMG is independent of thumb motion. We compared the accuracy and precision of AMG and EMG responses after patients' arms were immobilized during routine surgical positioning. METHODS: Patients undergoing general anaesthesia without neuromuscular block were enrolled. EMG and AMG monitors were placed on contralateral arms, and both arms were tucked under surgical drapes. The monitors simultaneously recorded train-of-four ratio (TOFR) after ulnar nerve stimulation every minute. Accuracy was defined as TOFR variability from the expected baseline value (1.00), and precision as consistency of repeated measurements, quantified by the coefficient of repeatability (CR). Failed measurements were defined as the absence of valid quantitative TOFR values, displayed as "0" or "?". AMG TOFR values were normalized to patient-specific baseline TOFR measurements for secondary analyses. RESULTS: Forty patients provided 3,733 TOFR measurements (1,879 EMG and 1,854 AMG). EMG was more accurate than AMG (mean EMG-AMG difference in absolute deviation from 1.0, -0.108; 95% confidence interval [CI], -0.134 to -0.083; p <  0.001) and more precise, with a smaller CR (0.10 vs. 0.35; p <  0.001), including compared with normalized AMG (CR, 0.35). Device failure rate was lower with EMG than AMG (3.6% vs. 27.7%; p =  0.0003). Failures increased over time, particularly for AMG after approximately 45 minutes. CONCLUSION: EMG provided more accurate and precise TOFR measurements than AMG, including normalized AMG, with fewer failures, supporting EMG-based monitoring when upper extremity mobility is restricted. REGISTRATION: Clinical trial registration: ClinicalTrials.gov Database (Registration Number: NCT06805032, Principal investigator: J.R.R., date of registration January 28, 2025).

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers