A patient contact shield visible in the localizer-the effects to tube current modulation.
Heli R S Larjava, Chibuzor Toivo Mark Eneh, Eini Niskanen, Lauri Järvinen, Joanna Sierpowska, Päivi Teerimäki, Jukka Järvinen, Virva Saunavaara,
Department of Medical Physics, Turku University Hospital and University of Turku, Kiinamyllynkatu 4-8, 20520 Turku, Finland. · Department of Radiology, Vaasa Central Hospital, Hietalahdenkatu 4, 65100 Vaasa, Finland.
Key points
From the paper's conclusion- Also, with some scanners, the use of XCare increased the cumulative tube current.
Abstract
An anthropomorphic phantom was scanned with Siemens Somatom computed tomography scanners using tube current modulation (TCM). Thorax of the phantom was first scanned without the use of a contact shield, followed by a scan with a 0.5 mm lead equivalent out-of-plane shield placed in the localizer but 1 cm caudally outside the scanned volume. Scans were repeated using the scanner's organ radiation dose reduction feature. Pitch values 0.35, 0.6, and 1.05 were used. With a pitch of 1.05, the effects of TCM varied between scanners from 4.35% to 356% when scanning toward the caudal end of the scanned volume, when shield was used outside the scanned volume. Also, with some scanners, the use of XCare increased the cumulative tube current.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- The impact of breast shielding on radiation dose during computed tomography procedures.Radiation protection dosimetry · 2026
- Tube current limitation in CT automatic exposure control across two technology eras: a 15-year retrospective comparison spanning scanner replacement and team-based protocol optimizationBMC Medical Imaging · 2026
- CBCT Dose Is Reduced with Angular Tube Current Modulation While Maintaining Image Quality and Reducing Photon Starvation Artifacts.Annals of biomedical engineering · 2026
- Exploring the effectiveness of CARE Dose 4D and CARE kV Automatic Exposure Control Systems in CT: A Slim and Obese Phantom StudyPolish Journal of Medical Physics and Engineering · 2026
- Influence of gantry overrun angle on CTDI variability: a Monte Carlo study.Radiological physics and technology · 2026
- Multi-system evaluation of baseline-referenced gray scale CNR profile deviation for phantom-based ultrasound quality assurance.Journal of applied clinical medical physics · 2026