Exploring the effectiveness of CARE Dose 4D and CARE kV Automatic Exposure Control Systems in CT: A Slim and Obese Phantom Study
Joanna Kidoń-Szołtysek, Aleksandra Klimas, Marcin Dybek, Leszek Wojciuch, Aneta Kawa-Iwanicka, Krzysztof S. Gołba
Key points
From the paper's conclusion- We focused on assessing how different parameter settings affect radiation dose and image quality.
Abstract
Introduction Radiology strives to minimise radiation exposure while maintaining diagnostic image quality. AEC systems achieve this by optimising radiation doses across different patients and anatomies. The purpose of our study was to evaluate the performance of AEC using Siemens CARE Dose4D and CARE kV options for both slim and obese adult phantoms. We focused on assessing how different parameter settings affect radiation dose and image quality. Material and methods Examinations were performed on a dual-source CT scanner using a slim and obese anthropomorphic phantom and a Catphan phantom. Parameters such as mAs modulation strength, reference kV values, quality reference mAs values, dose optimisation settings and CARE kV option were adjusted. The collected data included mA modulation, CTDI vol , effective mAs, quality reference mAs, reference mAs, kV values, and the relative standard deviation of HU values. Results The quality reference mAs serves as the baseline target for image quality and is influenced by the CARE kV mode, thereby designating the reference mAs (‘off ’ matches the baseline exactly; ‘on’ scales with the reference kV; ‘semi’ scales inversely with the reference kV). The effective mAs represents the final value influenced by all forms of modulation. For the slim phantom, the effective mAs remained lower than the reference mAs, while for the obese phantom, it was higher. The adjusted reference mAs values appear independent of phantom size. mA modulation curves reflect the link between CTD Ivol and AEC. Regardless of the phantom’s body type, CTDIvol decreased/increased when dose-saving optimiser levels, quality mAs ref , and kV ref were changed. Higher modulation strength lowers the dose for slim phantoms, but raises it for obese phantoms. Slim phantoms produce a lower relative standard deviation than obese ones. Conclusions Radiologists and medical physicists must understand the mechanics of the AEC system in CT scanners. This knowledge is essential for ensuring consistent image quality and minimising patient exposure.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- "Quantitative Modeling and Optimization of Radiation Dose in Routine CT Imaging: Balancing Image Quality and Dose Reduction under the ALARA Principle Framework"International Journal of Drug Delivery Technology · 2026
- Development of an Electronic Clinical Quality Measure for Excessive Radiation Dose or Inadequate Image Quality in Diagnostic CT in Adults.Journal of the American College of Radiology : JACR · 2026
- Continuous anatomical radiation modulation for the acquisition of coronary computed tomography angiography in difficult to scan patients: a prospective cohort study.Cardiovascular diagnosis and therapy · 2026
- Assessing variations in 3D image quality in chest CT across sites and scannersMedical Physics · 2026
- CT protocol optimisation for oncologic surveillance imaging: phantom-based comparison of multi-scan and single-scan acquisition strategies.Biomedical physics & engineering express · 2026
- Displayed CTDIvolume accuracy in ultra-low-dose CT: Limitations of current IEC criteria and a proposal for new constancy benchmarks.Radiography (London, England : 1995) · 2026