Evaluation of high-dose alerts in computed tomography using an automated dose monitoring system: a retrospective institutional study.
Mohammed Alanazi, Peter Kench, Seyedamir Tavakoli-Taba, Matthew Russell, Ernest Ekpo
Medical Image Optimisation and Perceptions Group, Discipline of Medical Imaging Science, Faculty of Medicine and Health, Susan Wakil Health Building, Camperdown Campus, Western Avenue, The University of Sydney, Camperdown, Sydney, NSW 2006, Australia.
Key points
From the paper's conclusion- Reasons for high-dose alerts included high body mass index (29%), scan repetition (25%), additional scans (15%), patient malpositioning (12%), excess scan range (11%), and orthopaedic hardware (6%).
Abstract
Radiation dose monitoring software (DMS) tools are increasingly used to support automated monitoring of radiation dose in medical imaging. This study evaluated the frequency and causes of high-dose events detected by a DMS in computed tomography (CT). All CT examinations performed at one hospital across three CT scanners over a one-year period were retrospectively analysed to establish the frequency and causes of high-dose alerts. Descriptive statistics summarized alert frequency and distribution, and multivariable logistic regression identified factors independently associated with high-dose alerts. Of 11,865 CT examinations, 851 (7%) exceeded predefined dose thresholds. High-dose alerts occurred most frequently in spine (17%), musculoskeletal joint (14%), abdomen/pelvis (8%), vascular angiography (8%), chest/lung (5%), and head/face (4%) examinations. Reasons for high-dose alerts included high body mass index (29%), scan repetition (25%), additional scans (15%), patient malpositioning (12%), excess scan range (11%), and orthopaedic hardware (6%). Multivariable analysis showed that overweight status, scanner model, spine and joint examinations, and patients aged 45-74 years were independently associated with high-dose alert occurrence (P < .05). DMS is an effective tool that enables imaging facilities to automatically identify high-dose events, facilitating investigation of their causes, protocol review, and identification of opportunities for dose optimisation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Early Implementation Performance of the CMS Electronic Clinical Quality Measure for Excessive Radiation Dose or Inadequate Image Quality in Diagnostic CT: An Evaluation of 254 Facilities.Journal of the American College of Radiology : JACR · 2026
- Development of an Electronic Clinical Quality Measure for Excessive Radiation Dose or Inadequate Image Quality in Diagnostic CT in Adults.Journal of the American College of Radiology : JACR · 2026
- Anatomy-Weighted CTDIvol from Routine CT Metadata: A Patient-Specific, Multi-Vendor Study Using Deep-Learning Segmentation.Tomography (Ann Arbor, Mich.) · 2026
- Assessment of CT doses and practices based on large data collected over four years in six hospitals of regional health network.Journal of medical imaging and radiation sciences · 2026
- Characterizing cumulative lens dose in recurrent CT: A Large-Scale Single-Center cohort study of drivers.European journal of radiology · 2026
- Leveraging longitudinal ACR CT Dose Index Registry data to improve interpretation of institutional CT dose performance.Journal of applied clinical medical physics · 2026