Early Implementation Performance of the CMS Electronic Clinical Quality Measure for Excessive Radiation Dose or Inadequate Image Quality in Diagnostic CT: An Evaluation of 254 Facilities.
Rebecca Smith-Bindman, Andrew B Bindman, Marc D Kohli, Carly Stewart, Timothy P Szczykutowicz, Ronald M Kline, Patrick S Romano
Department of Epidemiology and Biostatistics, University of California, San Francisco, San Francisco, California; Alara Imaging, Inc, Westport, Connecticut; Department of Obstetrics, Gynecology, and Reproductive Sciences, University of California, San Francisco, San Francisco, California; Philip R. Lee Institute for Health Policy Studies, University of California, San Francisco, San Francisco, CA. Electronic address: Rebecca@AlaraGateway.com. · Department of Epidemiology and Biostatistics, University of California, San Francisco, San Francisco, California; Department of Medicine, University of California, San Francisco, San Francisco, California, Emeritus.
Key points
From the paper's conclusion- Across 254 facilities and 783,634 examinations in 16 CT categories, 30.7% were out of range: 240,409 (30.7%) exceeded the dose threshold and 446 (0.06%) the noise threshold. Facility scores varied widely: median 29.5% (interquartile range [IQR] 18.4-44.7; range 1.8-82.4), with facilities at the 90th percentile scoring 4.6 times more out-of-range scans than those at the 10th percentile. Median out-of-range percentages by category ranged from 0% (routine and low-dose cardiac) to 82.6% (high-dose head; IQR 23.5-100.0); routine abdomen and pelvis, the highest-volume category, had a median of 43.9% (IQR 30.1-61.3). Within facilities, the median gap between best- and worst-performing category was 74 percentage points (IQR 49-89; intraclass correlation coefficient = 0.12).
Abstract
OBJECTIVE: To evaluate early facility performance on the CMS-adopted Excessive Radiation Dose or Inadequate Image Quality for Diagnostic CT in Adults (ExRad) measure, the first automated radiology outcome measure in CMS payment programs, during the first 16 months of the measure's inclusion in CMS programs. METHODS: ExRad evaluates each adult diagnostic CT examination in three steps: assignment to 1 of 19 CT categories defined by anatomic region and clinical indication; computation of size-adjusted radiation dose and image noise from the DICOM radiation dose-structured report and pixel data; and comparison of both against category-specific thresholds. Because the clinical and radiology data sit in separate systems, translation software inside the health system firewall may be used to perform the linkage. An examination is out of range if its dose or noise exceeds its category threshold; a facility's score is the percentage of eligible examinations out of range (lower is better). We analyzed all facilities reporting between May 2025 and August 2026. RESULTS: Across 254 facilities and 783,634 examinations in 16 CT categories, 30.7% were out of range: 240,409 (30.7%) exceeded the dose threshold and 446 (0.06%) the noise threshold. Facility scores varied widely: median 29.5% (interquartile range [IQR] 18.4-44.7; range 1.8-82.4), with facilities at the 90th percentile scoring 4.6 times more out-of-range scans than those at the 10th percentile. Median out-of-range percentages by category ranged from 0% (routine and low-dose cardiac) to 82.6% (high-dose head; IQR 23.5-100.0); routine abdomen and pelvis, the highest-volume category, had a median of 43.9% (IQR 30.1-61.3). Within facilities, the median gap between best- and worst-performing category was 74 percentage points (IQR 49-89; intraclass correlation coefficient = 0.12). DISCUSSION: Early ExRad performance varied widely across facilities and across CT categories within the same facility. Most variation reflected particular facility-category combinations, so facilities need their own category-level results to identify which protocols to target.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Development of an Electronic Clinical Quality Measure for Excessive Radiation Dose or Inadequate Image Quality in Diagnostic CT in Adults.Journal of the American College of Radiology : JACR · 2026
- Dose-aware 5-star image quality rating in CT: Findings from the IAEA-MGH study.European journal of radiology · 2026
- Evaluation of high-dose alerts in computed tomography using an automated dose monitoring system: a retrospective institutional study.Radiation protection dosimetry · 2026
- Leveraging longitudinal ACR CT Dose Index Registry data to improve interpretation of institutional CT dose performance.Journal of applied clinical medical physics · 2026
- Characterizing cumulative lens dose in recurrent CT: A Large-Scale Single-Center cohort study of drivers.European journal of radiology · 2026
- Identifying optimal dose head CT acquisition techniques through cluster analysis from 904,209 scans.European radiology · 2026