Neuro-Oncology

SURG-11 Comparison of Frame-Based and Frameless Robot-Assisted Biopsy

Iori Ozono, Fumiyuki Yamasaki, Ushio Yonezawa, Akira Taguchi, Yusuke Takeishi, Hiroki Taniguchi, Novita Ikbar Khairunnisa, Nobutaka Horie

Observational · n=147

Key points

From the paper's conclusion
  • Conclusion Frameless robot-assisted stereotactic biopsy using a twist-drill technique is a safe and effective alternative to conventional frame-based biopsy and significantly reduces postoperative intracranial air volume.

Abstract

Background Histopathological diagnosis remains essential for the management of intracranial lesions despite advances in neuroimaging. Although frame-based stereotactic biopsy has long been the standard approach, frameless robot-assisted systems have recently become available. This study compared conventional frame-based biopsy and frameless robot-assisted biopsy for supratentorial lesions. Methods We retrospectively reviewed 147 patients who underwent stereotactic biopsy for supratentorial lesions at our institution. Sixty-five patients underwent frame-based biopsy using the Leksell frame with a burr hole, and 82 underwent frameless robot-assisted biopsy using the Stealth Autoguide system with a twist-drill technique. Tissue sampling success, postoperative hemorrhage, and postoperative intracranial air volume were evaluated. Because the biopsy target is often difficult to identify on postoperative CT, intracranial air volume was used as a surrogate marker of targeting accuracy, based on previous reports demonstrating that intracranial air can cause brain shift and displacement of the target. Intracranial air volume was compared using the Mann–Whitney U test. In the frameless group, the relationship between sulcal width and postoperative intracranial air volume was also analyzed. Results Diagnostic tissue was successfully obtained in all cases. There was no significant difference in postoperative hemorrhage between the frame-based and frameless groups. The frameless group showed significantly lower postoperative intracranial air volume than the frame-based group (p = 0.0002). Within the frameless group, patients with narrower sulci at the target point demonstrated significantly lower postoperative intracranial air volume than those with normal or widened sulci (p < 0.0001). Conclusion Frameless robot-assisted stereotactic biopsy using a twist-drill technique is a safe and effective alternative to conventional frame-based biopsy and significantly reduces postoperative intracranial air volume.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

SURG-11 Comparison of Frame-Based and Frameless Robot-Assisted Biopsy — Science Skim