In-Flight Performance of Concentrator Versus Cylinder Oxygen During Mechanical Ventilation.
Mark L Munoz, Rittik Chaudhuri, Matthew J Eckert, Erik D Mueller
Medical Critical Care Service, Inova Fairfax Medical Campus, Falls Church, VA 22042-3300, United States. · Department of Emergency Medicine, Johns Hopkins University School of Medicine, Baltimore, MD 21287, United States.
Key points
From the paper's conclusion- Concentrator readings were generally lower; the median absolute difference was 3.15 pp, while upper-tail differences and configuration dependence preclude blanket interchangeability. The findings support configuration-specific evaluation of concentrators as cylinder extenders or candidate substitutes and evaluation of hybrid strategies in which compressed oxygen remains available for higher-demand conditions or contingencies. The study does not establish patient-level equivalence, isolated aircraft or ventilator effects, or universal replacement.
Abstract
INTRODUCTION: Reliable oxygen supply is a major challenge during prolonged casualty care and long-range aeromedical transport. Cylinders provide high-concentration oxygen but impose finite endurance, weight, high-pressure handling, and resupply requirements; concentrators can generate oxygen while power is available. To our knowledge, no prior peer-reviewed study has directly compared concentrator and cylinder oxygen during prolonged mechanical ventilation in actual fixed-wing flight. MATERIALS AND METHODS: Four flight exposures on a Dassault Falcon 50 and Bombardier Challenger 650 tested 4 complete concentrator-ventilator configurations. Oxygen flow 1-5 L/min was crossed with minute ventilation 5.0-15.0 L/min, yielding 100 matched active cells. Falcon source grids were paired across flights using ventilator-outlet readings; Challenger grids were paired within its 2-flight campaign using test-lung-entrance readings. Because the complete prespecified grid was observed, primary analyses were descriptive. We calculated signed and absolute source-difference summaries and displayed the complete grid. Exploratory factorial ordinary least squares models used categorical configuration, flow, minute ventilation, and flow-by-minute-ventilation terms with HC3 standard errors. The study evaluated equipment only and involved no humans or animals; institutional review board and animal care review were not applicable. RESULTS: The mean cylinder-minus-concentrator difference was 4.72 percentage points (pp; SD, 4.60); the concentrator was lower in 97 of 100 cells. The median absolute difference was 3.15 pp, and the 90th- and 95th-percentile absolute differences were 10.33 and 12.19 pp. Configuration means ranged from 1.39 to 7.96 pp. Sensitivity-analysis means ranged from 4.72 to 4.85 pp. Both aircraft powered the concentrator at 5 L/min directly from cabin outlets without modification or observed circuit interruption. Exploratory modeling identified configuration-associated differences. Tests of flow at minute ventilation 10 L/min, minute ventilation at flow 1 L/min, and the flow-by-minute-ventilation interaction were not significant. CONCLUSIONS: Concentrator readings were generally lower; the median absolute difference was 3.15 pp, while upper-tail differences and configuration dependence preclude blanket interchangeability. The findings support configuration-specific evaluation of concentrators as cylinder extenders or candidate substitutes and evaluation of hybrid strategies in which compressed oxygen remains available for higher-demand conditions or contingencies. The study does not establish patient-level equivalence, isolated aircraft or ventilator effects, or universal replacement.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Pilot implementation of 'outsourced oxygen to the bedside' models in five countries: a mixed methods impact assessment.BMJ public health · 2026
- Impact of oxygen delivery location during non-invasive home mechanical ventilation: a randomised cross-over simulation study.Respiration; international review of thoracic diseases · 2026
- External validation of predictive equations for determining the need for in-flight oxygen supplementation in patients with respiratory disease.Frontiers in physiology · 2026
- Oxygen therapy.Anales de pediatria · 2026
- Performance and Synchronization of Portable Oxygen Concentrators Delivering Pulsed Oxygen Therapy.Archivos de bronconeumologia · 2026
- Manual compared to automated closed-loop oxygen control strategies in hospitalised adults: a systematic review and meta-analysis.European respiratory review : an official journal of the European Respiratory Society · 2026