Impact of oxygen delivery location during non-invasive home mechanical ventilation: a randomised cross-over simulation study.

Dario Somaini, Zoe Bousraou, Lara Benning, Silvia Ulrich, Martin Lüchinger, Esther I Schwarz

Randomised

Key points

From the paper's conclusion
  • In this lung simulator study using a single-limb NIV circuit with a vented mask, oxygen injection location influenced the achieved FiO₂. Under the tested conditions, oxygen delivery via the ventilator oxygen port and a ventilator-near bleed-in connector resulted in broadly similar FiO₂ values and generally achieved higher FiO₂ than mask-near delivery. These findings warrant validation across different ventilators, circuits, interfaces, and pressure settings.

Abstract

INTRODUCTION: In non-invasive home mechanical ventilation (NIV) using single-limb circuits with vented masks, supplemental oxygen can be added at different locations, potentially resulting in variable delivered inspired oxygen fractions (FiO₂). Recommendations and data comparing oxygen inlet positions are lacking. METHODS: A lung simulator study was conducted using a 1.8 m single-limb NIV circuit with a vented mask. Oxygen was added at three locations: the ventilator oxygen port, a bleed-in connector near the ventilator outlet, and a bleed-in connector between the tube and mask. Oxygen flow rates of 1, 2, and 3 L/min were tested. Three simulated conditions were studied by adjusting lung compliance and airway resistance: healthy, COPD, and obesity. NIV was applied in pressure-controlled ventilation mode. Settings were EPAP/IPAP 4/14 cmH₂O for the healthy model, and both 4/18 and 8/20 cmH₂O for the COPD and obesity models, with a respiratory rate of 14/min. Mean resulting FiO₂ and minute ventilation from three repeated measurements in random order were recorded. RESULTS: Oxygen delivery via the ventilator oxygen-port and ventilator-near bleed-in produced broadly similar FiO2 values, ad both generally higher FiO₂ than mask-near delivery. Differences between inlet positions became more pronounced at higher oxygen flows reaching a maximum absolute difference of 5.3 percentage points between oxygen port an mask-near delivery. FiO₂ also differed between the two tested ventilator pressure settings in the COPD and obesity simulations, despite only small differences in minute ventilation. Increasing oxygen flow increased FiO₂ across all configurations, although the magnitude of this effect depended on oxygen-injection location and the simulated condition. CONCLUSION: In this lung simulator study using a single-limb NIV circuit with a vented mask, oxygen injection location influenced the achieved FiO₂. Under the tested conditions, oxygen delivery via the ventilator oxygen port and a ventilator-near bleed-in connector resulted in broadly similar FiO₂ values and generally achieved higher FiO₂ than mask-near delivery. These findings warrant validation across different ventilators, circuits, interfaces, and pressure settings.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers