Effect of sodium citrate and magnesium sulphate anticoagulation on platelet parameters measured on the Alinity hq analyzer.
Cristine Betzer, Rie McGrail
Department of Clinical Biochemistry, Gødstrup Hospital, Herning, Denmark.
Key points
From the paper's conclusion- Compared with EDTA, median platelet counts were statistically significantly lower in citrate (-21.1%) and MgSO4 (-31.7%) than in EDTA (P < .05). Mean platelet volume was also statistically significantly reduced in citrate (-3.2%) and MgSO4 (-4.7%) (P < .05). Median reticulated platelet fraction was slightly higher in both anticoagulants, but only MgSO4 differed substantially from EDTA.
Abstract
INTRODUCTION: EDTA-dependent pseudothrombocytopenia (PTP) may cause spuriously low platelet counts and can be prevented by using alternative anticoagulants, such as sodium citrate (citrate) and magnesium sulphate (MgSO4). However, these anticoagulants may affect platelet parameters depending on the analytical platform. Their effect on platelet measurements using the Abbott Alinity hq hematology analyzer has not previously been evaluated. METHODS: Blood samples from 40 outpatients without known PTP were collected in K2-EDTA, citrate, and MgSO4 tubes. Platelet count, mean platelet volume, and reticulated platelet fraction were measured on an Alinity hq analyzer. Paired comparisons were performed using Friedman repeated-measures analysis with the Dunn multiple comparisons test. RESULTS: Compared with EDTA, median platelet counts were statistically significantly lower in citrate (-21.1%) and MgSO4 (-31.7%) than in EDTA (P < .05). Mean platelet volume was also statistically significantly reduced in citrate (-3.2%) and MgSO4 (-4.7%) (P < .05). Median reticulated platelet fraction was slightly higher in both anticoagulants, but only MgSO4 differed substantially from EDTA. DISCUSSION: On the Alinity hq analyzer, citrate and MgSO4 result in statistically significantly lower platelet count than EDTA in samples from individuals without known PTP. Platelet parameters are therefore influenced by anticoagulant choice and should not be interpreted interchangeably.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- 405 The Effect of Anticoagulant Volume Adjustment in 3.2% Sodium Citrate Evacuated Tubes on the Viability of Prothrombin Time (PT), Partial Thromboplastin Time (PTT), and Fibrinogen Assay Among High Hematocrit Ethylenediaminetetraacetic Acid (EDTA) Tube Samples in the PhilippinesAmerican Journal of Clinical Pathology · 2026
- Impact of blood collection tubes on thrombin generation assay results: a comparison of citrate and CTADScientific Reports · 2026
- Concentration-dependent and platform-specific interference of argatroban and bivalirudin in routine coagulation assays.Thrombosis journal · 2026
- B-282 Analytical Validation and Comparison of Becton Dickinson and Greiner Bio-One Blood Collection TubesClinical Chemistry · 2026
- Factors contributing to thrombocytopenia after cardiac surgery; towards defining coagulation-friendly cardiac surgery.Perfusion · 2026
- Monitoring unfractionated heparin in ECMO and cardiac surgery: the search for a valid reference standardFrontiers in Medicine · 2026