Change management and continuous assurance of laboratory information system-based autoverification rules aligned with ISO 15189:2022: a practical framework for clinical laboratories.
Zilin Li, Lujuan Cui, Yinzhi Feng, Shiqian Feng
Department of Laboratory Medicine, Xichong County People's Hospital, Nanchong, Sichuan, China.
Key points
From the paper's conclusion- The framework consisted of a life cycle model; a taxonomy of change triggers; a 6-domain low, moderate, and high risk assessment; proportionate verification or revalidation methods; a minimum version control and deployment record; monitoring indicators; and discipline-specific considerations for chemistry, immunoassay, hematology, and coagulation. It also defined reference-adjudicated false-pass and false-hold measures and presented a wholly synthetic potassium example.
Abstract
INTRODUCTION: Laboratory information system (LIS)-based autoverification can improve turnaround time, consistency, and workload management, but rules may drift as analytical methods, reagent and calibrator lots, reference ranges, patient case mix, interfaces, and information systems change. This review proposes a standards-anchored framework for governing rules after rule implementation and maintaining patient safety across the rule life cycle. METHODS: Requirements from International Organization for Standardization (ISO) 15189:2022, ISO 22367:2026, Clinical and Laboratory Standards Institute (CLSI) AUTO15, archived CLSI AUTO10-A, and China Health Organization standard WS/T 616-2018 were integrated with select literature on autoverification and patient safety. No patient-identifiable data or new experimental data were used. RESULTS: The framework consisted of a life cycle model; a taxonomy of change triggers; a 6-domain low, moderate, and high risk assessment; proportionate verification or revalidation methods; a minimum version control and deployment record; monitoring indicators; and discipline-specific considerations for chemistry, immunoassay, hematology, and coagulation. It also defined reference-adjudicated false-pass and false-hold measures and presented a wholly synthetic potassium example. DISCUSSION: Laboratory information system-based autoverification rules should be managed as controlled, versioned, risk-sensitive clinical decision-support tools through structured change control, risk-proportionate verification or revalidation, and continuous monitoring throughout their life cycle.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- The effect of auto-verification on turnaround time for clinical chemistry and immunoassay tests.Journal of medical biochemistry · 2026
- A-266 Improving Laboratory Efficiency and Patient Safety Through Data-Driven Autoverification in Hematology and CoagulationClinical Chemistry · 2026
- Laboratory Information Management Systems and Quality Management: Development of a Framework and Future Directions for Histocompatibility and Immunogenetics Laboratories.International journal of immunogenetics · 2026
- From spreadsheets and scripts to controlled software lifecycles: recommendations for laboratory-developed software under ISO 15189 and IVDR.Clinical chemistry and laboratory medicine · 2026
- Validation of Methods in Medical Laboratories, ISO Standards, CLSI Guidelines, and Recommendations: A Point of ViewLaboratories · 2026
- B-050 Implementation and Performance Evaluation of Automated Autoverification for Tumor Marker Results in a High-Complexity Oncology LaboratoryClinical Chemistry · 2026