A patch-based bolus technique for total scalp irradiation: dosimetric impact of conformity and reproducibility.
Takahiro Kato, Hiroshi Niwayama, Yuki Tominaga, Yosuke Nagasawa, Daiki Senzaki, Takaomi Harada, Hisashi Sato
Department of Radiological Sciences, Graduate School of Health Sciences, Fukushima Medical University, 10-6 Sakaemachi, Fukushima, 960-8516, Japan. · Department of Radiology, Ohta Nishinouchi Hospital, 2-5-20 Nishinouchi, Koriyama, Fukushima, 963-8558, Japan.
Key points
From the paper's conclusion- These findings indicate that VMAT-based TSI is relatively robust against minor bolus variations.
Abstract
Total scalp irradiation (TSI) for scalp angiosarcoma remains technically challenging because of the large, highly curved target geometry and the need for reliable bolus application. This study aimed to develop a practical method to improve bolus conformity using commercially available bolus and to evaluate the dosimetric impact of bolus reproducibility in volumetric-modulated arc therapy (VMAT)-based TSI. An anthropomorphic head phantom was used to simulate TSI conditions. A three-segment 'patch bolus' method was developed to enhance conformity and setup stability. Treatment plans were generated using HyperArc with identical optimization parameters for three conditions: virtual bolus, single-sheet bolus and patch bolus. To assess reproducibility, the patch bolus was reapplied and rescanned 10 times, and dose distributions were recalculated on each dataset after rigid registration to isolate bolus-related variations. The patch bolus method substantially reduced air gaps compared with the single-sheet approach and achieved target coverage comparable to the virtual bolus condition. In contrast, the single-sheet bolus resulted in inferior coverage in regions with poor contact. Repeated application of the patch bolus demonstrated high reproducibility, with variations within 2% for most dose-volume metrics. Although relative variations exceeding 5% were observed for the lens, absolute doses remained low and clinically negligible. These findings indicate that VMAT-based TSI is relatively robust against minor bolus variations. The proposed patch bolus method provides improved conformity, reproducibility and workflow efficiency, representing a simple and clinically feasible strategy for TSI using external beam radiotherapy.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Updated dosimetric comparison of HDR interventional radiotherapy including 60Co sources versus VMAT and electron beam therapy for NMSC.Radiation and environmental biophysics · 2026
- Density override-based VMAT optimization to improve robustness at tissue-air interfaces in head and neck radiotherapy.Journal of applied clinical medical physics · 2026
- Dosimetric comparison of automated noncoplanar volumetric-modulated arc therapy and intensity-modulated proton therapy for total scalp irradiation in angiosarcoma of the scalp.The British journal of radiology · 2026
- Dosimetric Vulnerability of Radiotherapy Techniques Under Non-Rigid Arm-Position-Induced Deformations in Breast Cancer: A Deformable Phantom StudyRadiation · 2026
- Quantifying Patient Setup Error in Total Body Irradiation With Megavoltage Computed Tomography-guided Helical Tomotherapy.In vivo (Athens, Greece) · 2026
- ExacTrac Dynamic Surface-Guided Setup Versus Tattoo-Based Positioning in Breast Cancer Radiotherapy: A Retrospective Comparison of Setup Accuracy and Preparation Time.Cancers · 2026