Dynamic scan shaping: overcoming coil hysteresis for high-speed STEM
Jonathan J P Peters, Grigore Moldovan, Lewys Jones
Trinity College Dublin · Leibniz Institute of Plant Biochemistry
Key points
From the paper's conclusion- Here we demonstrate a predictive scan shaping approach using conventional scanning systems, where scan input is determined based on where the beam will be instead of where the beam should be.
Abstract
Scanning transmission electron microscopy (STEM) has a number of benefits over conventional parallel illumination transmission electron microscopy, such as the ability to simultaneously perform a range of imaging and spectroscopy techniques. However, one distinct disadvantage is the slow imaging speeds, a consequence of the sequential nature of the pixel acquisition, limiting imaging speeds to a few frames per second (fps). This reduces dose-rate control, increases the effect of distortions, and hinder the ability to capture dynamic events for in-situ experiments. The main obstacle to faster framerates is the inductance and hysteresis of the scanning coils that has previously only been addressable with the addition of entirely new hardware coils. Here we demonstrate a predictive scan shaping approach using conventional scanning systems, where scan input is determined based on where the beam will be instead of where the beam should be. We use this new approach to acquire fully sampled 512x512 images at 60 ns per pixel, giving a framerate of 41 fps.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Scan-coil delay causes anisotropic signal loss in fast 4D-STEM.Ultramicroscopy · 2026
- Autonomous Acquisition in Scanning Transmission Electron Microscopy for Dynamic Sampling of Material Heterogeneities.Microscopy and microanalysis : the official journal of Microscopy Society of America, Microbeam Analysis Society, Microscopical Society of Canada · 2026
- Self-supervised image denoising and restoration method for atomic force microscopy.Nature communications · 2026
- Adaptive Approach Strategy for Efficient Scanning Ion Conductance Microscopy Based on Non-monotonic Approach Curves.Analytical chemistry · 2026
- Drift correction of scan images by snapshot referencing.Ultramicroscopy · 2026
- Picometre-scale real-time drift correction in TEM and STEM by dynamic control of the specimen stage for atomic-resolution imaging.Ultramicroscopy · 2026