Scan-coil delay causes anisotropic signal loss in fast 4D-STEM.
Vishal Kumar, Andreas Jehle, Tizian Lorenzen, Julika Radecke, Knut Müller-Caspary, Massimo Kube, Henning Stahlberg
Laboratory of Biological Electron Microscopy, Institute of Physics, School of Basic Sciences, EPFL, Rt. de la Sorge, Lausanne, 1015, Switzerland; Department of Fundamental Microbiology, Faculty of Biology and Medicine, University of Lausanne, Rt. de la Sorge, Lausanne, 1015, Switzerland. Electronic address: Vishal.Kumar@epfl.ch. · Department of Chemistry and Center for NanoScience, Ludwig-Maximilians-Universität München, Butenandtstr. 11, München, 81377, Germany.
Key points
From the paper's conclusion- We present a phase-correlation-based sub-frame alignment procedure that measures and corrects this smearing, and we assess its impact on focused and defocused 4D-STEM reconstructions over a range of scan step sizes.
Abstract
Fast pixelated detectors are driving 4D-STEM toward microsecond dwell times, a regime in which the finite response of the scan deflection coils becomes comparable to the dwell time itself. Using direct probe imaging and sub-frame diffraction analysis, we document a significant intra-dwell scan-coil delay that systematically smears the recorded signal anisotropically along the fast scan direction, with a settling timescale of several tens of microseconds. We present a phase-correlation-based sub-frame alignment procedure that measures and corrects this smearing, and we assess its impact on focused and defocused 4D-STEM reconstructions over a range of scan step sizes. The correction restores signal across a broad range of spatial frequencies, with the largest gains at the large step sizes required for low-dose biological imaging. Because it operates on existing data with no modification to the microscope, the method offers a practical route to recovering signal that would otherwise be lost to scan-coil delay.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Dynamic scan shaping: overcoming coil hysteresis for high-speed STEMMicroscopy · 2026
- Pushing the dose limit of atomic-resolution imaging: A 4D-STEM case study of NaCl.Ultramicroscopy · 2026
- Interactive event-driven 4D-STEM with Timepix4 and live analytical ptychographyarXiv · 2026
- Universal Drift Correction for Multidimensional Scanning MicroscopyarXiv · 2026
- Drift correction of scan images by snapshot referencing.Ultramicroscopy · 2026
- Picometre-scale real-time drift correction in TEM and STEM by dynamic control of the specimen stage for atomic-resolution imaging.Ultramicroscopy · 2026