The journal of applied laboratory medicine

Revised Reference Intervals for Serum Free Light Chains: Implementation at the Mayo Clinic.

Maria Alice V Willrich, Jack L Wu, Brandon T Copet, Ashley R Thompson, Stephanie J Kalass, Carin Y Smith, Grant M Spears, Michael W Ettore,

Department of Laboratory Medicine and Pathology, Mayo Clinic, Rochester, MN, United States. · Division of Quantitative Health Sciences, Mayo Clinic, Rochester, MN, United States.

Key points

From the paper's conclusion
  • Legacy FLC RIs do not verify due to assay updates. Nonparametric 99% RIs better reflect contemporary assay performance and substantially reduce clinically meaningful misclassification.

Abstract

BACKGROUND: Universal reference intervals (RIs) for serum free light chains (FLCs) were established in 2002 using nephelometric instrumentation and early-generation reagents. Since then, assay platforms have evolved, renal effects have been characterized, and lot-to-lot drift has been documented. Large population-based studies also proposed age- and kidney function-stratified RIs. We evaluated whether legacy RIs remain appropriate following transition to a turbidimetric platform with stabilized reagents and established updated RIs. METHODS: Serum samples from 489 screened healthy adults were analyzed using the Optilite assay with stabilized κ calibrators. Samples with monoclonal proteins assayed by MASS-FIX, mass spectrometry-based monoclonal protein detection assay, were excluded. Central 99% RIs were derived using nonparametric methods. A secondary cohort of 106 individuals with chronic kidney disease (CKD) was analyzed to evaluate the effect of renal function. To assess clinical impact, 2025 clinical results, including >20 000 MASS-FIX-negative samples, were reclassified using legacy and revised RIs. RESULTS: The 2002 RIs were not verified in this contemporary cohort with 24.9% and 5.3% exceeding the historical upper limits for κFLC and κ/λ ratio, respectively. Newly established 99% RIs are 0.61 to 4.02 mg/dL for κ, 0.67 to 3.46 mg/dL for λ, and 0.57 to 2.45 for the κ/λ ratio. Lower estimated glomerular filtration rate (eGFR) was associated with increased FLC concentrations but smaller effects on the κ/λ ratio. Among MASS-FIX negative clinical samples, revised RIs reduced abnormal high flagging for κ/λ ratio from 17.6% to 4.1%. CONCLUSIONS: Legacy FLC RIs do not verify due to assay updates. Nonparametric 99% RIs better reflect contemporary assay performance and substantially reduce clinically meaningful misclassification.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers