European Heart Journal

Long-term incidence and efficacy of appropriate therapy in subcutaneous and transvenous implantable cardioverter-defibrillators: A secondary analysis of the PRAETORIAN-XL trial

Y Matsuda, L R A Olde Nordkamp, A Ghani, J Kuschyk, H Bonnemeier, M F El-Chami, A de Weger, L Smeding,

Trial · n=849

Key points

From the paper's conclusion
  • Purpose To compare the long-term incidence and efficacy of appropriate therapy in patients with S-ICD and TV-ICD, a secondary analysis of the PRAETORIAN-XL trial.

Abstract

Background Previous secondary analyses of the PRAETORIAN trial showed no differences in all appropriate therapy (shocks and antitachycardia pacing (ATP)) rates or first-shock efficacy between subcutaneous (S-ICD) and transvenous ICDs (TV-ICD) during 4.1 years of follow-up. However, ICD therapy is intended to provide lifelong protection, and data on the incidence and efficacy of appropriate therapy during long-term follow-up are lacking. The PRAETORIAN-XL trial extends follow-up by an additional 4 years to evaluate long-term appropriate therapy rates and first-shock efficacy. We hypothesize that S-ICD and TV-ICD demonstrate comparable performance during extended follow-up of more than 8 years. Purpose To compare the long-term incidence and efficacy of appropriate therapy in patients with S-ICD and TV-ICD, a secondary analysis of the PRAETORIAN-XL trial. Methods This is a secondary analysis of the PRAETORIAN-XL trial. From 39 centers in Europe and the United States, patients with a class I or IIa indication for implantable cardioverter-defibrillator (ICD) without the need for cardiac pacing were enrolled and randomized between March 2011 and January 2017. Patients were followed for an additional 4 years after the follow-up period of the PRAETORIAN trial. Appropriate therapy was defined as ATP or ICD shock for ventricular fibrillation (VF) or ventricular tachycardia (VT), and electrical storm was defined as 3 or more VF or VT episodes within 24 hours. Only appropriate therapies adjudicated with electrograms were analyzed, and analyses were performed according to a modified intention-to-treat principle. Results A total of 849 patients were randomized to the S-ICD group (n=426) and the TV-ICD group (n=423). During a median 7.3 (interquartile: 3.8-8.7) years follow-up, there was no difference in appropriate therapy rate between the S-ICD group and the TV-ICD group (116 and 103 patients, respectively; 8-year Kaplan-Meier estimated cumulative incidence: 31.7% vs. 29.8%; p=0.293) (Figure 1, 2). During follow-up, there is a linear annual increase in new patients receiving appropriate therapy (S-ICD: 2.6-8.6 per 100 patient-years, and TV-ICD: 3.1-6.8 per 100 patient-years). The first shock efficacy was similar between the S-ICD groups and the TV-ICD group (92.4% vs. 92.7%; p=0.888). The first ATP terminated 54.4% of monomorphic VTs (160/294), however, 9.9% of VT were accelerated and 0.7% of VT were converted to VF after ATP delivery (Figure 1). In total, 22 electrical storms occurred in 16 patients in the S-ICD group, and 24 electrical storms occurred in 20 patients in the TV-ICD group (p=0.482). Conclusions During long-term follow-up, there were no differences in the rate of appropriate therapy and first shock efficacy between S-ICD and TV-ICD. There is a linear increase in new patients receiving appropriate therapy over the years, emphasizing the long-term benefit of ICD therapy with the current guidelines.Figure 1.The details of the therapiesFigure 2.The rate of appropriate therapy

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers