European Heart Journal

ECG-based automated 3D localization of earliest right ventricular activation

L Ding, M Tang, S J Zhou

Observational

Key points

From the paper's conclusion
  • Overall error was 7.44±2.88 mm for AAOL-3D vs 6.80±3.46 mm for AAOL-Surface (P < 0.001).

Abstract

Background The Automated Arrhythmia Origin Localization (AAOL) algorithm predicts the location of earliest ventricular activation in real time using standard ECG and patient-specific electroanatomic mapping (EAM) geometry. We previously extended AAOL to output 3D coordinates (AAOL-3D) in a multicenter cohort; however, right-ventricular (RV) evaluation was limited by a small sample size. Objective To validate AAOL-3D for localizing RV pacing sites (earliest activation locations) in a larger RV cohort and to compare its performance with AAOL-Surface. Methods RV pacing data from patients with premature ventricular contraction were analyzed. For each site, 120-ms QRS integrals (leads III, V2, V6) and 3D EAM coordinates were extracted. Within each patient, leave-one-site-out cross-validation was performed (test site; remaining sites for training). AAOL-Surface predicted a point on the EAM surface; AAOL-3D predicted full 3D coordinates. Localization error was the Euclidean distance between predicted and true coordinates. Paired tests compared methods overall and by RV subregion (RVOT vs non-RVOT). Results A total of 20 patients with 196 sites were enrolled (9.8±1.7 sites/patient). The mean age was 47.1 years and 63.6 % were female. The common comorbidities including hypertension (25.0%), diabetes (9.4%), and heart failure (6.3 %). Overall error was 7.44±2.88 mm for AAOL-3D vs 6.80±3.46 mm for AAOL-Surface (P < 0.001). RVOT (n=57): 7.83±2.92 vs 7.39±3.48 mm (P=0.133). Non-RVOT (n=139): 7.28±2.86 vs 6.55±3.43 mm (P < 0.001). Figure 1 shows statistical results and a pacing-site localization example. Conclusion Both methods achieved clinically acceptable accuracy (<10 mm). In RVOT, AAOL-3D was comparable to AAOL-Surface, supporting feasibility of direct 3D coordinate prediction in this anatomically complex, crescent-shaped region where surface projection can introduce geometric artifacts. Further evaluation and optimization across RV subregions are indicated.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

ECG-based automated 3D localization of earliest right ventricular activation — Science Skim