A faster, optimized pulsed field timing sequence demonstrates safety under aggressive lesion stacking without evidence of electrode heating
A Amin, B Gidney, J Romero, T Gomez, E Byun, Q Chen, C Beeckler, C Porterfield,
Key points
From the paper's conclusion- These findings suggest that the accelerated pulse protocol preserves PFA’s non-thermal safety characteristics and remains well-tolerated even during excessive use conditions.Figure 1
Abstract
Background The Variable Loop Circular Catheter (VLCC) is a pulsed-field ablation (PFA) device for treatment of atrial fibrillation (AF) that has demonstrated a favorable safety and efficacy profile in clinical use. A novel pulse protocol, Sequence 2, which shortens ablation delivery from the ~20.6 second pulse delivery of Sequence 1 to ~3.8 seconds, is untested for its ability to be safe under excessive use conditions. Because the shorter delivery time increases instantaneous pulse density, it is pertinent to assess if Sequence 2 creates a risk of focal electrode heating or energy stacking during repeated applications. Purpose The objective of this study was to assess the safety and efficacy of the new VLCC Sequence 2 pulse recipe, when used outside the bounds of the recommended clinical workflow. Methods A total of eight porcine subjects were randomized into two test groups: 2X nominal dose and 4X nominal dose, corresponding to two or four consecutive PFA applications per site, respectively. PFA was performed with intentional lesion stacking and nominal electrode irrigation (30ml/min). Subsequent coronary angiography, blood chemistry panels, and evaluation of VLCC electrode charring was also performed. Animals were survived for 30±3 days post ablation to undergo follow-up remapping procedures before being euthanized to allow for gross necropsy and histopathology. Results For all animals, 100% acute and chronic effectiveness was observed for pulmonary vein isolation (PVI). Safety metrics were met, no incidence of phrenic nerve injury, esophageal injury, or catheter charring were observed. Transient asymptomatic coronary spasm could be induced and was detected following cavotricuspid ablation in both the 2X and 4X dose groups, whereas spasm following mitral isthmus ablation only occurred in the 2X group. All coronary spasms resolved following administration of a nitroglycerin dose and no hemodynamic compromise was observed. The 4X group exhibited greater transient increases in hemolysis (e.g., plasma-free hemoglobin, LDH) and myocardial injury markers (e.g., CK/troponins) compared with the 2X group, with all values returning to or near baseline by the survival endpoint. The 4X treatment group showed more marked changes in hemolysis and myocardial injury markers than the 2X treatment group, with all markers returning to or near baseline by the survival endpoint (Figure 1). Conclusion In this study, Sequence 2 maintained a favorable safety profile even when applied at 2X and 4X nominal dosing with no evidence of charring or excessive electrode heating at nominal irrigation and only transient elevations in hemolysis and myocardial injury markers. PFA applications near coronary arteries, under these excessive use conditions, resulted in self-limiting spasm. These findings suggest that the accelerated pulse protocol preserves PFA’s non-thermal safety characteristics and remains well-tolerated even during excessive use conditions.Figure 1
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Effect of Pulsed Field Timing Sequence and Electrode Cooling on Ablation Safety and Efficacy Surrogates: A Preclinical Study.Circulation. Arrhythmia and electrophysiology · 2026
- Timing of consecutive ablations during PFA: implications of back-to-back or interval spaced lesion stackingEuropean Heart Journal · 2026
- Electrolytic Electroporation (E2): first in vivo demonstration for cardiac ablationEuropean Heart Journal · 2026
- Practical Guide on the Use of the PulseSelect Pulsed Field Ablation System for Atrial Fibrillation Ablation Procedures.Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2026
- PVI Using the Fixed Loop PFA in Paroxysmal AF: One-Year Outcomes with Continuous Implantable Loop Recorder Monitoring.Indian pacing and electrophysiology journal · 2026
- Asymmetric biphasic pulsed field ablation of paroxysmal atrial fibrillation: clinical assessment.Frontiers in cardiovascular medicine · 2026