Safety and effectiveness of a mid-voltage pulsed-field ablation catheter in persistent atrial fibrillation: acute outcomes

B Ren, R Zeng, X Hu, F Li, W Li, A Gong, G Lv, Y Du,

Observational · n=127

Key points

From the paper's conclusion
  • Purpose To evaluate the acute safety and effectiveness of a small-balloon, mid-voltage PFA catheter for pulmonary vein isolation (PVI) and linear ablation in patients with persistent atrial fibrillation (PersAF).

Abstract

Background High-voltage pulse field ablation (PFA) can disrupt red blood cell membranes, leading to an increased risk of hemolysis. A novel small-balloon, mid-voltage PFA system is expected to reduce the risk of hemolysis through lower pulse voltages while achieving favorable acute success rates. Purpose To evaluate the acute safety and effectiveness of a small-balloon, mid-voltage PFA catheter for pulmonary vein isolation (PVI) and linear ablation in patients with persistent atrial fibrillation (PersAF). Methods A prospective, multicenter, single-arm study was conducted at seven centers on 127 patients with drug-refractory PersAF. All patients underwent PVI using the small-balloon PFA catheter integrated with a mid-voltage PFA generator (850 V), and an electroanatomic mapping system (Panel A-B). In addition to PVI, investigators could choose one or more linear ablation strategies, including the left-atrial roof or floor, mitral isthmus, or right-atrial tricuspid isthmus. Endpoints were the incidence of post-procedural serious adverse events within 7-days, laboratory parameters related to hemolysis to assess asymptomatic hemolysis and procedural effectiveness within acute phase. Results Patients’ mean age was 60.53 ± 8.30 years (73.2% male), with a median of 152 [114-217] applications per patient. Acute success of PVI and linear-block was 100% (127/127) and 99.2% (126/127), respectively. Mean procedure, fluoroscopy, and ablation times were 126.5 ± 32.8, 10.5 ± 5.7, and 30.4 ± 14.5 min, respectively. Changes from baseline in haptoglobin, indirect bilirubin, and lactate dehydrogenase 2-days post-procedure were -0.10 ± 0.12 g/L, 7.55 ± 7.52 μmol/L, and 98.95 ± 58.72 U/L, respectively (Panel C). No serious adverse events related to delivery of PFA using the integrated system occurred within the 7-day perioperative period. Conclusion This clinical trial demonstrated that the small-balloon, mid-voltage PFA delivery and mapping system safely achieved superior acute PVI and linear ablation in patients with PersAF. The use of mid-voltage PFA may explain the reduced risk of hemolysis. Twelve-month outcomes will be reported when available.Panel

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Safety and effectiveness of a mid-voltage pulsed-field ablation catheter in persistent atrial fibrillation: acute outcomes — Science Skim