Association between HALT incidence and deformation of transcatheter aortic valve prostheses: Sapien vs. Evolut vs. Navitor
K Onishi, M Yasuda, N Soejima, A Yoshida, K Fujita, H Matsuzoe, T Miyoshi, M Ueno,
Key points
From the paper's conclusion- Methods We retrospectively evaluated 125 consecutive patients (Sapien: n = 54, Evolut: n = 39, Navitor: n = 32) who underwent transcatheter aortic valve replacement from December 2022 to April 2024.
Abstract
Background Hypoattenuating leaflet thickening (HALT) has been associated with the eccentricity of implanted transcatheter aortic valve (TAV). However, its relationship with valve type (balloon-expandable vs. self-expandable) or valve design (supra-annular vs. intra-annular) remains unclear. The purpose of our study was to investigate the incidence of HALT and valve eccentricity among TAV series (Sapien, Evolut, and Navitor). Methods We retrospectively evaluated 125 consecutive patients (Sapien: n = 54, Evolut: n = 39, Navitor: n = 32) who underwent transcatheter aortic valve replacement from December 2022 to April 2024. HALT and eccentricity of the implanted TAV were analyzed with 30-day postprocedural computed tomography. HALT was defined as involvement of more than 25% from the leaflet base in at least one leaflet, based on the Valve Academic Research Consortium 3 criteria. Results The overall incidence of HALT was 30.4%, with a significantly higher incidence in Navitor than in Sapien or Evolut (56.2% vs. 22.2% vs. 20.5%, p = 0.001). In all three groups (Sapien, Evolut, and Navitor), eccentricity at the leaflet inflow level was significantly greater in patients with HALT compared to without HALT (median 0.38 [IQR 0.36-0.45] vs. 0.34 [0.29-0.39]; p = 0.040, 0.59 [0.39-0.67] vs 0.36 [0.32-0.41]; p < 0.001, 0.64 [0.57-0.77] vs 0.44 [0.40-0.51]; p < 0.001). Eccentricity at the native annulus level was significantly lower in Sapien compared with Evolut and Navitor (0.36 [0.33-0.42] vs. 0.60 [0.48-0.67] and 0.59 [0.49-0.70], respectively; p < 0.001). Eccentricity at the leaflet inflow level was significantly greater in Navitor compared with Sapien and Evolut (0.36 [0.31-0.39] and 0.36 [0.33-0.44] vs. 0.57 [0.45-0.67], respectively; p < 0.001). Conclusion The self-expandable, intra-annular design is more affected by the device landing zone and results in leaflet level deformation, leading to increased incidence of HALT.For image description, please refer to the figure legend and surrounding text.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Early hemodynamic impact of hypoattenuated leaflet thickening in new-generation transcatheter heart valvesEuropean Heart Journal · 2026
- Incidence, predictors, and clinical impact of hypoattenuating leaflet thickening following SAPIEN 3 Ultra RESILIA implantationEuroIntervention · 2025
- Clinical significance of CT-detected hypoattenuated leaflet thickening after transcatheter aortic valve replacement: A systematic review and meta-analysis.Journal of cardiovascular computed tomography · 2026
- Case Report: Serial 4D-CT evolution from focal leaflet thickening to multileaflet HALT after transcatheter aortic valve replacement in a hemodialysis patient.Frontiers in cardiovascular medicine · 2026
- Post-TAVR Neo-Sinus Geometry and Impaired Ventricular Recovery Are Associated with Early Leaflet Thrombosis: First MDCT Insights from Vietnam.Journal of clinical medicine · 2026
- Real-Time Fluoroscopic Assessment of Underexpansion Predicts Hemodynamic Valve Deterioration Following TAVR With Balloon-Expandable DeviceJACC: Cardiovascular Interventions · 2025