European Heart Journal

Wearable 12-lead electrocardiogram for detecting acute myocardial ischemia: a prospective diagnostic accuracy study

Y Wang, B Y Yu, C C Tam, H F Tse

Observational · n=503

Key points

From the paper's conclusion
  • Conclusions The wearable 12-lead ECG provides high diagnostic accuracy for detecting ST-segment deviations and performs comparably to the standard 12-lead ECG in identifying OMI and localizing the culprit vessel, supporting its potential as a reliable and accessible tool for early triage in acute chest pain.

Abstract

Background Timely and accurate identification of myocardial ischemia is critical for improving outcomes in patients with acute chest pain. Although the conventional 12-lead electrocardiogram (ECG) is the standard for detecting ST-segment changes, its dependence on specialized equipment may delay diagnosis. Wearable ECG devices offer a more accessible alternative, but their diagnostic accuracy remains inadequately validated. Purpose To assess the diagnostic accuracy of a wearable 12-lead ECG in detecting ST-segment deviations and occlusive myocardial infarction (OMI) in patients presenting with chest pain. Methods This prospective study enrolled patients presenting with acute chest pain to the emergency department of a tertiary hospital between May 2025 and January 2026. Both conventional and wearable 12-lead ECGs were recorded prior to coronary intervention. Two senior cardiologists, blinded to clinical and angiographic outcomes, independently interpreted all ECGs. OMI was defined as a culprit lesion with Thrombolysis in Myocardial Infarction (TIMI) flow grade 0–2 or TIMI flow grade 3 accompanied by a peak high-sensitivity troponin T level of >1000 ng/L. Results A total of 503 patients were enrolled (mean age 66.4±15.1 years; 66.7% male). Using the conventional 12-lead ECG as the reference standard, the wearable ECG demonstrated high diagnostic accuracy for ST-segment elevation (sensitivity 92.7%, specificity 99.4%) and depression (sensitivity 91.8%, specificity 98.9%). Among 57 patients with confirmed OMI, the wearable ECG showed comparable sensitivity (75.4% vs 78.9%, P=0.48) and specificity (91.2% vs 89.7%, P=0.56) to the conventional 12-lead ECG. The accuracy of culprit vessel identification by ECG was also similar between the two modalities (63.2% vs 66.7%, P=0.52). Conclusions The wearable 12-lead ECG provides high diagnostic accuracy for detecting ST-segment deviations and performs comparably to the standard 12-lead ECG in identifying OMI and localizing the culprit vessel, supporting its potential as a reliable and accessible tool for early triage in acute chest pain.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers