Evaluation of a Wrist-Worn Device for ECG-Based Detection of Acute Myocardial Infarction.
Karolina Eitminaviciene, Daivaras Sokas, Zygimantas Abramikas, Dziugile Kersnauskaite, Dominika Sirviene, Artiomas Sirvys, Guoste Stankeviciute, Vytautas Juknevicius,
Biomedical Engineering InstituteKaunas University of Technology Kaunas 44249 Lithuania. · Institute of Clinical Medicine, Faculty of MedicineVilnius University Vilnius 01513 Lithuania.
Key points
From the paper's conclusion- The two-lead ECG acquired with a wrist-worn device offers diagnostic value and enables screening for AMI.
Abstract
OBJECTIVE: To determine whether a two-lead ECG (wECG), acquired using a self-applicable, wrist-worn device, provides diagnostic value in acute myocardial infarction (AMI) detection. METHODS: Seventy-one patients with AMI, 26 with other cardiovascular diseases (CVDs), and 54 healthy controls were enrolled in the study. To acquire lead I, one electrode of the wrist-worn device was touched with the index finger of the opposite hand, while another lead was acquired by touching the electrode on the strap to a specific part of the body. The diagnostic value of the wECG is evaluated either directly or through synthesis of the standard 12-lead ECG using an echo state network. RESULTS: Using the standard ECG, an AMI detector based on a convolutional neural network yields a sensitivity of 0.86 and a specificity of 0.71; the corresponding figures for blinded cardiologist diagnosis are 0.77/0.88. Using the synthesized ECG, the V3 electrode touch site yields the best performance, with a sensitivity/specificity of 0.84/0.71 for AMI detection and 0.67/0.86 for cardiologist diagnosis. Compared to the synthesized ECG across all touch sites, use of the wECG results in a substantially lower sensitivity, reduced by 0.12-0.13, while the specificity is comparable or slightly higher. Implementation of a two-stage approach, consisting of detector-based screening followed by cardiologist review of positive cases, results in correct decisions in 63% of the AMI cases using either the synthesized ECG or the wECG. CONCLUSION: The two-lead ECG acquired with a wrist-worn device offers diagnostic value and enables screening for AMI.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Wearable 12-lead electrocardiogram for detecting acute myocardial ischemia: a prospective diagnostic accuracy studyEuropean Heart Journal · 2026
- Artificial intelligence–based ECG as a triage tool for acute myocardial infarction: a diagnostic systematic review and meta-analysisEuropean Heart Journal · 2026
- Assessing the diagnostic accuracy of wearable ECG monitors in detecting ischaemiaEuropean Heart Journal · 2026
- Artificial intelligence-based ECG as a triage tool for acute myocardial infarction: a diagnostic systematic review and meta-analysis.European heart journal. Digital health · 2026
- From wrist to heart: clinical validation and real-world translation of wearable and portable AI-enhanced electrocardiography.European heart journal. Digital health · 2026
- Diagnostic Accuracy of Artificial Intelligence for Detecting ST-Elevation Myocardial Infarction on 12-Lead Electrocardiograms: A Systematic Review and Meta-Analysis.Cardiology in review · 2026