When Left Is Right: Multimodality Insights Into How Left Internal Jugular Access Optimizes Device Trajectory and Geometry in Transcatheter Tricuspid Valve Replacement
Kevin Lai, Georgi K Fram, John Dawdy, James Alter, Hani Hamade, Bryan M Zweig, Sachin Parikh, Arturo Giacaman Saavedra,
Key points
From the paper's conclusion- Conclusions In this single-centre experience, LIJ access facilitated more central placement of EVOQUE delivery system in the septal-lateral tricuspid annular plane and improved trajectory alignment in TTVR with EVOQUE which correlated with comparable procedure times and safety compared to other access choices.
Abstract
Background Edwards EVOQUE system is a transcatheter tricuspid valve replacement (TTVR) option designed for right transfemoral access, however, suboptimal valve positioning and delivery occurs in patients with unfavorable anatomy. When considering alternative access, early uses of left internal jugular (LIJ) access has seen procedural success and hosts advantageous mechanisms that are poorly understood. Aims To understand the mechanism of LIJ approach by developing a mechanistic framework that demonstrates the geometric advantages compared to right internal jugular (RIJ) and how it can improve EVOQUE valve positioning and delivery Methods and results Retrospective analysis of 24 patients (LIJ = 13, RIJ = 11) who underwent successful TTVR procedure between June 2024 and September 2025. Multimodality analysis combining echocardiography, cardiac computed tomography and fluoroscopy was performed to analyze delivery system positions LIJ demonstrated larger delivery system distance to septal wall (16.6 ± 5.3 mm) than RIJ (9.5 ± 3.5 mm)(P < 0.001) and was similarly observed at 10 mm and 20 mm proximal points, across 3 time points as validated by transesophageal echocardiography septal distances to delivery system. Trajectory of delivery system averaged LIJ = 17.8 ± 4.8° vs RIJ = 2.3 ± 2.8° (P < 0.010). LIJ primary flexion angle on average was larger at 82.0 ± 14.1° vs 57.2 ± 21.8°, P = 0.003. CT 3D modeling estimated larger average distance of delivery system to septal plane for LIJ (24.4 ± 9.1 mm vs 19.3 ± 7.3 mm) and length of primary flexion to tricuspid annulus (52.4 ± 8.0 mm vs 44.4 ± 5.7mm, P < 0.05). Conclusions In this single-centre experience, LIJ access facilitated more central placement of EVOQUE delivery system in the septal-lateral tricuspid annular plane and improved trajectory alignment in TTVR with EVOQUE which correlated with comparable procedure times and safety compared to other access choices. While these findings offer a mechanistic rationale for LIJ access in selected patients, larger prospective multi-centre studies are required to confirm clinical superiority and refine patient selection criteria.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Device-lead abnormalities and function after transcatheter tricuspid valve replacementEP Europace · 2025
- Interventional Tricuspid Valve Replacement: A Structured Sequence of Procedures.Journal of visualized experiments : JoVE · 2026
- Implantation Depth and New Conduction Abnormalities After Transcatheter Tricuspid Valve Replacement.Structural heart : the journal of the Heart Team · 2026
- Transjugular LuX-Valve Plus transcatheter tricuspid valve replacement for torrential tricuspid regurgitation in a patient with a giant right atrium and multiple prior valve interventions: a case report.European heart journal. Case reports · 2026
- Right internal jugular approach for transcatheter edge-to-edge mitral valve repair using the MitraClip G4 system after failed transfemoral access: a case report.European heart journal. Case reports · 2026
- The TriArtic procedure: TRIcuspid annular reduction by TEER in the commissure-a transcatheter annuloplasty enabling EVOQUE tricuspid valve replacement: a case report.European heart journal. Case reports · 2026