The TriArtic procedure: TRIcuspid annular reduction by TEER in the commissure-a transcatheter annuloplasty enabling EVOQUE tricuspid valve replacement: a case report.
Raphael Phinicarides, Fabian Voß, Tobias Zeus, Amin Polzin
Department of Cardiology, Pulmonology, and Vascular Medicine, University Hospital Düsseldorf, Medical Faculty of the Heinrich Heine University Düsseldorf, 40225 Düsseldorf, Germany.
Key points
From the paper's conclusion- We performed the TriArtic procedure—a single TriClip XTW (Abbott) deployed in the septo-anterior position as a transcatheter annuloplasty step.
Abstract
BACKGROUND: Transcatheter tricuspid valve replacement (TTVR) with the EVOQUE system has expanded therapeutic options for severe tricuspid regurgitation (TR), but a prohibitively dilated tricuspid annulus may exceed the device-sizing matrix and preclude implantation. No transcatheter strategy has been described to mechanically pre-condition such an annulus. CASE SUMMARY: A 71-year-old woman with heart failure with reduced ejection fraction (left ventricular ejection fraction 38%, suspected premature-ventricular-complex-induced cardiomyopathy), permanent atrial fibrillation, and class III obesity presented with torrential atrial functional TR refractory to fully optimized guideline-directed medical therapy. Screening cardiac computed tomography (CT) classified the native annulus as 'outside parameters' for the 56 mm EVOQUE (Edwards Lifesciences); the Heart Team also deemed standalone tricuspid edge-to-edge repair (T-TEER) inadequate owing to the massively dilated annulus. We performed the TriArtic procedure-a single TriClip XTW (Abbott) deployed in the septo-anterior position as a transcatheter annuloplasty step. Repeat CT after 27 days demonstrated an ∼11% reduction in tricuspid annular area and conversion of projected-derived-diameter oversizing from -2.9% to +3.3% (within parameters). Transfemoral EVOQUE 56 mm was implanted 48 days after the TriClip; postprocedural imaging confirmed no residual regurgitation and a mean transtricuspid gradient of 1 mm Hg, with rapid recovery of cardiorenal function. DISCUSSION: Transcatheter tricuspid annuloplasty by a strategically placed commissural TEER device may convert a prohibitively dilated annulus into one within EVOQUE-implantable parameters. In carefully selected patients ineligible for either standalone strategy, sequential transcatheter annuloplasty followed by valve replacement-the TriArtic procedure-may offer a feasible pathway, conceptually analogous to the mitral ARCTIC technique. This single case demonstrates feasibility only and does not establish general applicability.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- ANTecedent Annular Reduction to Facilitate TransCatheter TrICuspid Valve Replacement (ANTARCTIC)JACC Case Reports · 2026
- Novel Transcatheter Tricuspid Valve Replacement Device for Recurrent Regurgitation After Failed Surgical Repair.JACC. Case reports · 2026
- The TRICURE FIH study results and future trials updateEuropean Heart Journal · 2026
- Transcatheter tricuspid valve replacement-a comparison of clinical and procedural outcomes of two different prostheses.Frontiers in cardiovascular medicine · 2026
- Interventional Tricuspid Valve Replacement: A Structured Sequence of Procedures.Journal of visualized experiments : JoVE · 2026
- Orthotopic Transcatheter Tricuspid Valve Replacement After Heterotopic Caval Valve Failure.JACC. Case reports · 2026