Implantation Depth and New Conduction Abnormalities After Transcatheter Tricuspid Valve Replacement.
Amir Ghabousian, Anas I Zaqut, Ashtyn Chamberland, Zachary Taylor, Howard K Song, Megan McGregor, Firas E Zahr, Scott M Chadderdon
Division of Cardiovascular Medicine, Oregon Health and Science University, Portland, Oregon, USA.
Key points
From the paper's conclusion- Approximately 1 in 5 patients undergoing EVOQUE TTVR developed an ONP, which was independently associated with a shallower septal anchor implantation depth. This study results are hypothesis-generating and requires validation in larger, prospective cohort studies.
Abstract
BACKGROUND: Transcatheter tricuspid valve replacement (TTVR) has been shown to markedly improve patients' quality of life compared to medical therapy. Unfortunately, a potential consequence of this TTVR is the development of new conduction abnormalities requiring permanent pacemaker implantation. As such, we examined the association between septal anchor implantation depth and the development of clinically significant conduction abnormalities following EVOQUE TTVR. METHODS: We conducted a single-center, retrospective review of all EVOQUE TTVR cases from May 2020 to May 2025. Anterior-, mid-, and posterior-septal anchor depths were measured relative to the tricuspid annular plane and the average depth of the septal anchors was calculated. The primary outcome was obligate need for pacing (ONP), consisting of new permanent pacemaker implantation or new pacemaker-dependence within 30 days. RESULTS: Within 30 days of TTVR, 9 of 47 patients (19.1%) developed ONP. The mean septal depth was significantly closer to the tricuspid annulus in patients with ONP compared with those without ONP (2.40 ± 0.70 mm vs. 3.75 ± 1.37 mm; p < 0.001). This shallow septal depth was associated with a higher risk of ONP within 30 days in univariable analysis and remained independently associated in a multivariate analysis after adjustment for prior surgical aortic valve replacement (adjusted odds ratio per 1-mm increase: 0.44; 95% CI: 0.20-0.96; p = 0.038). CONCLUSIONS: Approximately 1 in 5 patients undergoing EVOQUE TTVR developed an ONP, which was independently associated with a shallower septal anchor implantation depth. This study results are hypothesis-generating and requires validation in larger, prospective cohort studies.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Leadless Pacing for Delayed Heart Block After Transcatheter Tricupid Valve Replacement.JACC. Case reports · 2026
- Device-lead abnormalities and function after transcatheter tricuspid valve replacementEP Europace · 2025
- Transcatheter Tricuspid Valve Replacement in High-Risk Elderly Patients: Procedural Feasibility, Complexity, and Delayed Conduction Disturbances.JACC. Case reports · 2026
- Safety and Feasibility of Left Bundle Branch Area Pacing Following Transcatheter Tricuspid Valve ReplacementJACC Case Reports · 2026
- Incidence and Predictors of Acute Tricuspid Regurgitation Worsening After Left Bundle Branch Area Pacing.Circulation reports · 2026
- Impact of basal septal bulge on conduction disturbance relative to valve position after transcatheter aortic valve replacementEuropean Heart Journal · 2026