Key points

From the paper's conclusion
  • We report the case of a 77-year-old woman with a degenerated 23-mm Mitroflow valve (Sorin Group Inc) and severe aortic regurgitation who was deemed extreme risk for redo surgery.

Abstract

Coronary artery obstruction is a rare but life-threatening complication of valve-in-valve transcatheter aortic valve replacement, particularly in patients with externally mounted leaflet bioprostheses and high-risk computed tomography features. We report the case of a 77-year-old woman with a degenerated 23-mm Mitroflow valve (Sorin Group Inc) and severe aortic regurgitation who was deemed extreme risk for redo surgery. Because BASILICA was not feasible due to heavily calcified leaflets, valve-in-valve transcatheter aortic valve replacement with a 20-mm SAPIEN 3 Ultra (Edwards Lifesciences) was performed with simultaneous bilateral coronary chimney stent deployment (TRITON technique). The procedure was successful, with preserved coronary flow, excellent valve function, and an uncomplicated clinical course at 6-month follow-up. This case illustrates a feasible rescue/preventive strategy for selected patients at prohibitive risk of coronary obstruction when leaflet modification is not possible.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers