B-340 Gel You Later: a real-world workflow evaluation of the EXENT® System versus electrophoretic methods
Steven Conklin-Lopez, Elizabeth Crogan, Mitzi Scales, Eric Roberts, Nicholas Paciaroni
Key points
From the paper's conclusion- These findings indicate that automated mass spectrometry-based workflows may meaningfully reduce labor burden in the clinical laboratory.
Abstract
Background Serum protein electrophoresis (SPE) with reflex immunofixation electrophoresis (IFE) remains an integral component for diagnosis and monitoring of monoclonal gammopathies by detecting and characterizing monoclonal proteins (MPs). However, conventional gel-based SPE/IFE workflows are labor-intensive, highly manual, subject to interpretive variability, resulting in substantial technologist hands-on time. The FDA-cleared EXENT® System, an automated MALDI-ToF mass spectrometry-based platform, was recently developed to streamline MP identification and characterization with reduced operator intervention. This study quantitatively compared workflow demands between conventional gel-based SPE/IFE testing and the EXENT® system using real-world laboratory data. Methods A time-and-motion workflow study was conducted at Labcorp (Burlington, NC). Hands-on and walk-away times were recorded for operators performing SPE and IFE on traditional electrophoretic systems and on the automated EXENT® system (Thermo Fisher Scientific). One instrument was used for SPE and six for IFE. For the SPE/IFE workflows, observations included one SPE operator, two IFE operators, and one dedicated interpretation operator. For the EXENT® workflow, a single instrument was utilized by two operators who participated in the study; however, the workflow was performed by only one operator at a time. This included a new user who completed the manufacturer’s standard 2.5-day hands-on training. Batch level processing times were evaluated for the analysis (SPE/IFE: n=90 batches; EXENT®: n=50 batches). Recorded time during the study was segregated between hands-on time (sample preparation, interpretation and maintenance) and walk-away time. EXENT® workflow times reflect averages across seven independent runs. Results EXENT® operators independently completed the full workflow, processing 150 samples (three 50-sample batches) within an eight-hour shift, including six hours of walk-away time. The new operator achieved full proficiency shortly after training and reproduced the manufacturer’s throughput claims. On a per-sample basis, EXENT® demonstrated a 56% reduction in total hands-on time relative to the SPE/IFE workflow, including a 73% reduction in interpretation time, a 67% reduction in maintenance time, and a 125% increase in walk-away time. Additional workflow details are provided in the supporting figure. Conclusion The EXENT® system was easy to learn, reliable in routine use, and rapidly incorporated into a high-volume clinical workflow. Labcorp operators confirmed that a single operator could process 150 samples in an eight-hour shift, consistent with manufacturer claims. Compared with conventional gel-based SPE/IFE workflows, EXENT® markedly reduced operator hands-on time, particularly for interpretation, and significantly increased walk-away time. These findings indicate that automated mass spectrometry-based workflows may meaningfully reduce labor burden in the clinical laboratory.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- B-039 Harmonizing M-protein analysis: an international, multi-center reproducibility study of the EXENT® System and evaluation of concordance to electrophoretic methodsClinical Chemistry · 2026
- B-196 Comprehensive clinical validation of the Immunoglobulin Isotypes (GAM) assay for the EXENT® Analyzer assay in monoclonal gammopathiesClinical Chemistry · 2026
- Performance evaluation of an automated mass spectrometry system for the measurement and isotyping of monoclonal immunoglobulinsClinica Chimica Acta · 2026
- B-064 Bridging Quantitative MALDI-TOF Mass Spectrometry to IMWG Complete Response: Deriving a Serum Cut-off Equivalent to Negative ImmunofixationClinical Chemistry · 2026
- Harmonizing M-protein analysis: an international, multi-center reproducibility study of mass spectrometry and evaluation of concordance to electrophoretic methods.Clinical chemistry and laboratory medicine · 2026
- Automated peak detection, isotyping and quantitation of M-proteins by MALDI-TOF MS for replacement of immunofixation and serum protein electrophoresisJournal of Immunological Methods · 2026