Clinical Chemistry

B-277 Analytical Comparison of ANA Screening by BioPlex® 2200 and the Sebia dIFine® IFA Platform

Isabel Espinosa, Mark Pool, Merih Tesfazghi

Key points

From the paper's conclusion
  • Methods Serum samples (n = 101) submitted to the core laboratory for ANA screening as part of routine patient care were sequentially analyzed using the Bio-Rad BioPlex 2200 and the Sebia dIFine® IFA platforms according to the manufacturers’ instructions.

Abstract

Background Autoimmune connective tissue diseases (CTDs) are common and often present with heterogeneous or overlapping clinical manifestations, making early and analytically robust laboratory detection essential to prevent irreversible organ damage. Many patients exhibit polyautoimmunity or evolving autoantibody profiles, with sequential development of additional specificities over time. This evolution increases the risk that restricted screening strategies may fail to detect concurrent or emerging autoimmune conditions, potentially delaying appropriate management. Indirect immunofluorescence assay (IFA) on HEp-2 cells remains the reference method for antinuclear antibody (ANA) screening because of its broad antigenic substrate and ability to detect antibodies against a wide spectrum of nuclear and cytoplasmic targets. In this study, we evaluated the diagnostic performance of ANA screening using the Sebia automated IFA platforms (dIFine and P30 with Sebia HEp-2 slides), with the in-house BioPlex 2200 multiplex assay serving as the predicate method. Methods Serum samples (n = 101) submitted to the core laboratory for ANA screening as part of routine patient care were sequentially analyzed using the Bio-Rad BioPlex 2200 and the Sebia dIFine® IFA platforms according to the manufacturers’ instructions. The dIFine automated imaging software was used to capture and classify IFA images, while manual IFA was performed to titer positive results identified by the Bio-Rad BioPlex 2200. When sufficient sample volume was available, discrepant results were further evaluated by manual IFA using Sebia HEp-2 slides and interpreted by an experienced technologist. Positive and negative percent agreement, as well as Cohen’s kappa (MedCalc online), were calculated to assess method concordance. Results Overall concordance between the two methods was 75% (76/101), with a Cohen’s kappa of 0.42 (95% CI, 0.23–0.61), indicating moderate agreement. Negative percent agreement was 80.3% (57/71), while positive percent agreement was 63.3% (19/30). Fourteen discrepant samples were negative by the Bio-Rad BioPlex 2200 but positive by the Sebia dIFine system; 10 of these were confirmed by manual IFA, with titers ranging from 1:40 to 1:640. Conclusion Given the limited sample size, the two methods demonstrated overall moderate agreement, with high negative percent agreement. Additional studies are ongoing to expand the sample size and incorporate a third method to aid in discrepancy resolution and more accurately characterize diagnostic performance.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers