B-187 Clinical Performance Comparison of Manual, Semi-Automated, and Fully Automated Anti-Nuclear Antibody Immunofluorescent Platforms for Screening Systemic Autoimmune Rheumatic Diseases
Camille Novis, Tammy Trudeau, Mark Parish, Toshia Chavez, Vijayalakshmi (Viji) Nandakumar
Key points
From the paper's conclusion- These findings support the clinical interchangeability of ANA IFA platforms while highlighting that platform-specific specificity differences, particularly with NOVA View, may inform laboratory implementation decisions.
Abstract
Background Anti-nuclear antibody (ANA) testing by indirect immunofluorescence assay (IFA) remains the gold standard screening method for systemic autoimmune rheumatic diseases (SARDs). With increasing adoption of automated ANA platforms, careful evaluation of clinical performance is necessary to ensure reliable screening and consistent interpretation across laboratory settings. This study evaluated and compared the clinical performance of three ANA IFA platforms representing manual, semi-automated, and fully automated workflows to guide implementation decision in our clinical laboratory. Methods Residual sera from 210 subjects undergoing routine clinician-ordered ANA testing and 40 healthy controls were analyzed using three ANA IFA systems. The manual workflow used the Kallestad™ HEp-2 kit (Bio-Rad, CA, USA) interpreted by fluorescence microscopy (Olympus CX43, INOVA Diagnostics, USA). The semi-automated workflow employed the NOVALite® HEp-2 ANA IFA kit (Inova Diagnostics, USA), with slide processing on the QUANTA-Lyser 3000 and digital image–guided interpretation using the NOVA View® system. The fully automated workflow utilized the Helios™ Automated IFA System (AESKU.GROUP, Wendelsheim, Germany), providing automated slide processing, digital image acquisition, and software-assisted interpretation of AESKUSLIDES® ANA HEp-2 slides. Final pattern and titer determinations were performed by a trained laboratory technologist. Clinical performance was assessed at a positive screening cutoff of 1:80, with performance metrics presented alongside 95% confidence intervals in brackets. Statistical analyses were conducted using Analyse-it and GraphPad Prism software. Results At the 1:80 dilution, the manual Bio-Rad system demonstrated sensitivity of 98% (87–100%) and specificity of 49% (42–56%). The fully automated Helios system showed sensitivity of 95% (84–99%) and specificity of 53% (46–60%), while the semi-automated NOVA View system demonstrated sensitivity of 90% (77–96%) and specificity of 58% (51–65%). Across platforms, positive predictive values ranged from 30% to 32% (23–41%) and negative predictive values from 96% to 99% (91–100%), consistent with ANA’s role as a sensitive screening assay. Positive likelihood ratios ranged from 1.91 to 2.15 (1.65–2.62) and negative likelihood ratios from 0.05 to 0.17 (0.01–0.43), with odds ratios between 12.81 and 38.32 (4.38–284.56). Cohen’s kappa values ranged from 0.25 to 0.28, and Youden’s index from 0.46 to 0.48. When analyzed separately, specificity varied substantially among healthy controls, at 65% for Bio-Rad, 80% for Helios, and 98% for NOVA View. In the disease cohort, specificity ranged from 45% to 47% across platforms. Combined disease and healthy control specificity ranged from 49% to 58%. Receiver operating characteristic analysis demonstrated comparable discrimination across platforms, with area under the curve values of 0.80 (p < 0.0001). Conclusion All three ANA IFA platforms demonstrated similar overall clinical performance, characterized by high sensitivity and negative predictive value at a 1:80 cutoff, with modest specificity and low positive predictive value. Notably, the NOVA View platform demonstrated substantially higher specificity in healthy controls and the highest combined specificity overall, suggesting potential advantages in reducing false-positive results while maintaining acceptable sensitivity. These findings support the clinical interchangeability of ANA IFA platforms while highlighting that platform-specific specificity differences, particularly with NOVA View, may inform laboratory implementation decisions.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- B-277 Analytical Comparison of ANA Screening by BioPlex® 2200 and the Sebia dIFine® IFA PlatformClinical Chemistry · 2026
- Is there an alternative to the indirect immunofluorescence ANA HEp-2 assay for the diagnosis of connective tissue diseases?Frontiers in Immunology · 2025
- B-140 Performance and Workflow Impact of Automated Image Analysis in High-Volume ANA HEp-2 IIF ScreeningClinical Chemistry · 2026
- Real-World Evaluation of ANA Testing Pathways: Serological Incompleteness and Comparison of Stepwise Versus Upfront Parallel Reflex Strategies.Biomedicines · 2026
- B-178 Comparative Performance of ANCA Indirect Immunofluorescence Assays and Concordance of Manual Versus Automated Interpretation in 583 SeraClinical Chemistry · 2026
- B-141 Antinuclear Antibody (ANA) HEp-2 Immunofluorescence Pattern Distribution in a Large Brazilian Laboratory Cohort: A Ten-Year Retrospective StudyClinical Chemistry · 2026