B-178 Comparative Performance of ANCA Indirect Immunofluorescence Assays and Concordance of Manual Versus Automated Interpretation in 583 Sera
D O N G L A I MA, Danielle Accardi, John LaPira, Kathleen Cichonski, Julius Torres, Alberto Torregroza, Mark Little, Matthieu Coq,
Key points
From the paper's conclusion- We compared two commercial assays and assessed manual to automated concordance within one platform.
Abstract
Background Variation across assays and readout modes may affect standardization of ANCA indirect immunofluorescence (IIF) in ANCA associated vasculitis (AAV). We compared two commercial assays and assessed manual to automated concordance within one platform. Methods 583 clinically characterized sera collected at the Trinity College Dublin Kidney Center were tested (223 AAV: 123 microscopic polyangiitis, 92 granulomatosis with polyangiitis, 8 eosinophilic granulomatosis with polyangiitis; 360 disease controls) on ethanol and formalin fixed human granulocyte substrates from two manufacturers (Sebia; Werfen NOVA Lite) using a 1:20 cut off per instructions for use. Manual reading was used for cross assay comparison. Within the Sebia assay, manual reading was compared with a computer-assisted image interpretation approach using the dIFine® automated microscope. Outcomes were sensitivity/specificity (manual) and positive/negative/overall agreement (manual vs automated). Results With manual reading, ethanol fixed slides showed sensitivity of 96.0% (Sebia) vs 92.8% (NOVA Lite) and specificity of 62.5% vs 61.4%. Formalin fixed slides showed sensitivity of 87.4% vs 88.8% and specificity of 92.5% vs 90.6%. Within Sebia, the manual vs automated analysis agreement yielded 96.6%/96.6%/96.6% (ethanol) and 93.7%/99.8%/97.4% (formalin). Conclusion In this single center cohort, the two assays demonstrated comparable manual performance. Automated reading achieved high agreement with manual interpretation within the tested Sebia assays, supporting feasibility for standardized ANCA IIF workflows. Multi-center studies should evaluate reproducibility and workflow impact.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- ANCA testing in real-world clinical practice: diagnostic performance and predictive value in a Spanish cohortLara D. Veeken · 2026
- ANCA practice in diagnostic pathology: state of play and unresolved challenges.Rheumatology (Oxford, England) · 2026
- Is there an alternative to the indirect immunofluorescence ANA HEp-2 assay for the diagnosis of connective tissue diseases?Frontiers in Immunology · 2025
- B-187 Clinical Performance Comparison of Manual, Semi-Automated, and Fully Automated Anti-Nuclear Antibody Immunofluorescent Platforms for Screening Systemic Autoimmune Rheumatic DiseasesClinical Chemistry · 2026
- B-140 Performance and Workflow Impact of Automated Image Analysis in High-Volume ANA HEp-2 IIF ScreeningClinical Chemistry · 2026
- Antineutrophil Cytoplasmic Antibodies: Integrating History, Pathogenesis, and Modern Diagnostics.International journal of rheumatic diseases · 2026