B-174 Method Comparison and Validation of Rubella IgG Testing on the DiaSorin Liaison XL

Alice Ann Lever, Susan Healy, Yun (Wayne) Wang

Key points

From the paper's conclusion
  • Conclusion Diasorin Liaison XL Rubella IgG assay is acceptable for implementation.

Abstract

Background IgG serology testing for the Rubella Virus is utilized for establishing immunity status in patients, particularly in prenatal care and employee health screening. The immunology laboratory at an urban academic medical center is undergoing a modernization project, including switching Rubella IgG testing from the Beckman DxI 800 to the Diasorin Liason XL. The aim of this study is to validate the Diasorin Liason XL Rubella IgG assay and ensure there is agreement with the current DxI 800 assay results. Methods A ten-point simple precision study with low positive and positive controls was used to confirm performance within manufacturer specifications. Forty-four known patient samples were run on both machines to confirm accuracy using qualitative comparison (positive, equivocal, or negative). Specificity was verified according to the manufacturer’s instructions for use. Some discrepancies were identified during method comparison, especially differences in reporting units and assay standardization. To further evaluate performance at the cutoff range, the DxI 800 control material was serially diluted in saline and analyzed four times to assess qualitative agreement relative to expected IU/mL values. Results Initial method comparison showed a qualitative agreement of 81.8%, with discordant results including 7 samples classified as equivocal by the DxI 800 and negative by the Liaison XL, and one sample classified as negative by the DxI 800 and positive by the Liaison XL. Differences in reporting units and cutoff thresholds were found between the two machines. The DxI 800 reports results in IU/mL and is standardized to the WHO 2nd International Standard (positive cutoff 15.0 IU/mL), whereas the Liaison XL reports index values corresponding to a positive cutoff equivalent to 10 IU/mL under the WHO 1st International Standard. Serial dilution of DxI control material resulted in 100% agreement following adjustment for cutoff equivalency. The low positive control showed a mean of 1.497 index with 1.4% CV, and the positive control showed a mean of 13.07 index with 1.8% CV. Both were within manufacturer precision performance claims, which were deemed acceptable. Conclusion Diasorin Liaison XL Rubella IgG assay is acceptable for implementation. Initial discordance between the DxI 800 and the Liason XL was attributed to differences in reporting units and WHO standard calibration. Recognition of these differences and using serial dilutions to determine accuracy resulted in agreement between methods.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

B-174 Method Comparison and Validation of Rubella IgG Testing on the DiaSorin Liaison XL — Science Skim