  • These findings support a tiered testing approach: sensitive assays like ELiA and ELISA can identify potential SLE cases, with results interpreted alongside ANA positivity and confirmed with specific methods like CLIFT.

Abstract

Background Anti–double-stranded DNA (dsDNA) antibodies are a key biomarker for systemic lupus erythematosus (SLE) and are included in the 2019 American College of Rheumatology classification criteria. In clinical practice, dsDNA testing supports SLE diagnosis following a positive ANA (anti-nuclear antibody) screen and compatible clinical features. Assay performance varies due to differences in antigen substrates, assay technologies, and antibody avidity detection. This study evaluates the analytical and clinical performance of three commonly used dsDNA assays—enzyme-labelled anti-isotype assay (ELiA), enzyme-linked immunosorbent assay (ELISA), and Crithidia luciliae indirect immunofluorescence test (CLIFT)—using an automated fluorescence platform. Methods A total of 218 specimens submitted for concurrent ANA IFA (immunofluorescence assay) and anti-dsDNA testing were analyzed, along with 40 healthy donors. Clinical history was available for 104 patients, classified as SLE (n=32) or non-SLE controls (n=72) based on physician diagnosis. Testing included ANA IFA on Kallestad HEp-2 Complete Kit (Bio-Rad, CA, USA) with manual microscopy (Olympus CX43, Inova Diagnostics, USA); Phadia ELiA dsDNA Ab on the Phadia 250 (Thermo Fisher Scientific, Uppsala Sweden); QUANTA Lite dsDNA ELISA (Inova Diagnostics, USA); and CLIFT (NOVA Lite DAPI Crithidia luciliae kit, USA), processed on the QUANTA-Lyser 3000 and read using the NOVA View imager. Assays were performed per manufacturers’ instructions, with positivity defined by stated cutoffs. Clinical performance metrics are presented with 95% confidence intervals in brackets. Analyses were conducted using Microsoft Excel and Analyse-it. Results Using ANA IFA as reference, ELiA showed PPA 74% (66–80%), NPA 57% (48–65%); ELISA showed PPA 40% (33–49%), NPA 88% (81–92%); and CLIFT showed PPA 26% (20–34%), NPA 99% (96–100%). Clinical performance metrics—sensitivity, specificity, PPV, NPV, accuracy, LR+, and LR–—were, respectively: for ELiA, 100% (89–100%), 1% (0–7%), 31% (23–41%), 100% (21–100%), 32% (24–41%), 1.01(0.99–1.04), and 0.00 (0.00–0.24); for ELISA, 72% (55–84%), 74% (62–82%), 55% (40–69%), 85% (75–92%), 73% (64–81%), 2.72 (1.75–4.24), and 0.38 (0.22–0.68); and for CLIFT, 59% (42–74%), 94% (87–98%), 83% (63–93%), 84% (74–90%), 84% (75–90%), 10.69 (3.95–28.89), and 0.43 (0.28–0.66). ROC analysis showed AUCs of 0.805 (ELISA), 0.777 (CLIFT), and 0.772 (ELiA) (p<0.0001). At 95% specificity optimized cut-off, ELiA demonstrated sensitivity of 47% (31-64%), ELISA 44% (28-61%), and CLIFT 56% (39-73%), with PPV values of 83% (61-94%), 82% (59-94%), and 86% (65-95%), respectively. Conclusion ELiA, ELISA, and CLIFT exhibit distinct analytical and clinical profiles in detecting anti-dsDNA antibodies. ELiA’s high sensitivity is offset by extremely low specificity, and a high false-positive rate, which leads to unnecessary reflex testing. ELISA offers balanced sensitivity and specificity, while CLIFT provides the highest specificity and PPV, supporting its role as a confirmatory assay despite lower sensitivity. Optimizing cut-offs improves predictive values and likelihood ratios across assays. These findings support a tiered testing approach: sensitive assays like ELiA and ELISA can identify potential SLE cases, with results interpreted alongside ANA positivity and confirmed with specific methods like CLIFT.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers