Clinical Chemistry

B-157 Analytical evaluation of a fully automated multiplex microarray immunoassay for simultaneous detection of anti-MPO, anti-PR3, and anti-GBM autoantibodies in a reference laboratory in Southern France

Gerber Gomez, Guillaume Nardella, Claire Monat

Key points

From the paper's conclusion
  • This study evaluated the analytical performance of a fully automated multiplex microarray immunoassay (MosaiQ AiPlex® VAS; AliveDx, Switzerland; investigational device) for simultaneous detection of IgG antibodies to MPO, PR3, and GBM, compared with a routinely used method in a reference clinical laboratory in Southern France.

Abstract

Background Accurate detection of disease associated autoantibodies is essential for evaluating anti-glomerular basement membrane (GBM) disease and ANCA-associated vasculitides (AAV). Clinical assessment often requires measurement of various serological targets, including antibodies to myeloperoxidase (MPO), proteinase 3 (PR3), and GBM. This study evaluated the analytical performance of a fully automated multiplex microarray immunoassay (MosaiQ AiPlex® VAS; AliveDx, Switzerland; investigational device) for simultaneous detection of IgG antibodies to MPO, PR3, and GBM, compared with a routinely used method in a reference clinical laboratory in Southern France. Methods De-identified, remnant human sera with known MPO, and GBM IgG antibody status were tested using the multiplex microarray assay. Results were compared with the laboratory’s routine method, an addressable laser bead immunoassay (ALBIA). Agreement was assessed using positive percent agreement (PPA), negative percent agreement (NPA), overall percent agreement (OPA), and Cohen’s kappa. Immunofluorescence (IF) results were reviewed for discrepant samples when available but were not included in the primary analysis (Table 1). Results A total of 387 results were included, comprising 100 antibody positive samples (anti-MPO: 52; anti-PR3: 48; anti-GBM: 0) and 287 antibody negative samples (anti-MPO: 122; anti-PR3: 126; anti-GBM: 39). For anti-GBM, comparison was limited to negative samples because no positive specimens were available. Across markers, the investigational device showed high concordance with the comparator, with an overall percent agreement of 97.9% and a Cohen’s kappa of 0.94. PPA and NPA were:Anti-MPO: 92% and 99.2% (excluding equivocal results)Anti-PR3: 95.8% and 99.2%Anti-GBM: not applicable and 100% Three equivocal results (all anti-MPO) were observed (1.7% of MPO positive samples; 0.8% overall). When treated as positive, anti-MPO’s PPA and NPA were 92.3% and 98.4%; when treated as negative, 88.5% and 99.2%. No invalid results occurred. Additional details are summarized in Table 1. Discordant result review:Anti-MPO: Four samples were negative on the investigational device but positive on ALBIA. All showed a perinuclear-ANCA pattern by IF, with titers from borderline (1:20) in two samples, to strongly positive (1:160–1:320) in the remaining two. One additional sample was borderline positive on the investigational device but negative on ALBIA.Anti-PR3: Two samples were negative on the investigational device but positive on ALBIA. One showed a high titer (1:320) cytoplasmic-ANCA pattern by IF. The other had a low positive ALBIA value, a perinuclear-ANCA IF pattern (1:320), and concurrent MPO positivity, on both compared platforms suggesting a potential false positive PR3 result on ALBIA. Finally, one sample was positive on the investigational device but negative on ALBIA; it showed a high titer (1:320) cytoplasmic-ANCA IF pattern, consistent with a possible false negative PR3 result on the comparator. Interpretation of discordant findings was limited by available supporting data. Conclusion The multiplex microarray immunoassay demonstrated strong analytical agreement with the routine ALBIA comparator for detecting antibodies to MPO, and GBM. These results support the suitability of this fully automated, high throughput platform as a consolidated testing approach in the laboratory evaluation of anti-GBM disease and ANCA-associated vasculitides.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

B-157 Analytical evaluation of a fully automated multiplex microarray immunoassay for simultaneous detection of anti-MPO, anti-PR3, and anti-GBM autoantibodies in a reference laboratory in Southern France — Science Skim