B-328 MALDI-TOF Mass Spectrometry Demonstrates Superior Analytical Specificity Over Serum Immunofixation for Anti-CD38 Therapeutic Background
Cristina Agullo Roca, Sergio Castro, Itxaso Jimenez, Verónica González-Calle, Elena Alejo, Borja Puertas, Beatriz Rey, María Isidoro-García,
Key points
From the paper's conclusion- We therefore evaluated the frequency of theoretical and clinically relevant overlap between endogenous IgGk/k MPs and anti-CD38 mAbs and compared analytical interference rates between MS and sIFE in patients receiving daratumumab or isatuximab.
Abstract
Background Anti-CD38 monoclonal antibodies (mAbs), including daratumumab and isatuximab, are widely used in multiple myeloma (MM) and may interfere with serum immunofixation electrophoresis (sIFE) due to shared IgGk isotype and electrophoretic co-migration with endogenous monoclonal proteins (MPs). As use of these therapies increases, laboratory interference has become a significant analytical concern. MALDI-TOF mass spectrometry (MS) discriminates immunoglobulins based on precise mass-to-charge (m/z) values rather than migration patterns. Therapeutic antibodies have defined m/z signatures (daratumumab 11,691; isatuximab 11,745), and potential overlap is limited to narrow ±4 Da windows and their +75 Da matrix adduct peaks, including the ±4 Da window around the adduct position. We therefore evaluated the frequency of theoretical and clinically relevant overlap between endogenous IgGk/k MPs and anti-CD38 mAbs and compared analytical interference rates between MS and sIFE in patients receiving daratumumab or isatuximab. Methods We retrospectively analyzed 1101 serum samples using the EXENT® system (The Binding Site, part of Thermo Fisher Scientific) and sIFE by SAS Vitrési system (Helena BioSciences). Of these, 505 patients had IgGk or k light-chain MPs and were therefore at theoretical risk of m/z overlap with anti-CD38 IgGk mAbs. Within this subgroup, theoretical overlap was defined as ±4 Da coincidence with the mAb m/z or its +75 Da adduct. Among these 505 IgGk/k patients, 125 were actively receiving anti-CD38 therapy (daratumumab or isatuximab) and were evaluated for clinically relevant analytical interference. Interference rates between MS and sIFE were compared using Fisher’s exact test. Results Among 505 patients with IgGk or k light-chain MPs, theoretical m/z proximity within ±4 Da was observed in 6.34% (32/505) for daratumumab and 3.76% (19/505) for isatuximab, accounting for both direct peak coincidence and potential overlap with the +75 Da matrix adduct peak. However, when restricted to 125 patients receiving anti-CD38 therapy, clinically relevant spectral overlap by MS was identified in only 2 cases (1.9%), whereas sIFE demonstrated interference in 43 cases (34%). This corresponds to 31.5-fold higher odds of interference with sIFE compared to MS (OR 31.5; p<0.0001) and a 95% relative reduction in observed therapeutic antibody interference using MS. The markedly lower interference rate observed with MS reflects its intrinsic analytical specificity. Spectral overlap is structurally constrained to two narrowly defined ±4 Da mass windows corresponding to the therapeutic antibody m/z or its +75 Da matrix adduct peak; outside these discrete regions, mass resolution preserves clear discrimination between endogenous monoclonal proteins and therapeutic antibodies. In contrast, sIFE relies on electrophoretic co-migration, a broader separation mechanism that permits substantially higher rates of apparent overlap and reduced analytical specificity. Future incorporation of free light-chain–specific immunoenrichment beads may further mitigate rare IgG-based overlap scenarios by MS by enabling independent confirmation of endogenous k light-chain signals. Conclusion These findings demonstrate that while theoretical spectral overlap may occur in a minority of IgGk cases, true analytical interference using MS is rare and substantially lower than that observed with electrophoretic methods. Implementation of MS-based monitoring significantly reduces false-positive therapeutic antibody interference and improves analytical accuracy in MM response assessment.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Resolving Therapeutic Monoclonal Antibody Interference in an Immunoglobulin M Kappa Multiple Myeloma Patient Using a Mass Spectrometry Assay.Clinical laboratory · 2026
- B-064 Bridging Quantitative MALDI-TOF Mass Spectrometry to IMWG Complete Response: Deriving a Serum Cut-off Equivalent to Negative ImmunofixationClinical Chemistry · 2026
- B-065 Retrospective MALDI-TOF Mass Spectrometry Anticipates Relapse and Resolves Therapeutic Antibody Interference in IgD/L Multiple MyelomaClinical Chemistry · 2026
- Performance evaluation of an automated mass spectrometry system for the measurement and isotyping of monoclonal immunoglobulinsClinica Chimica Acta · 2026
- B-196 Comprehensive clinical validation of the Immunoglobulin Isotypes (GAM) assay for the EXENT® Analyzer assay in monoclonal gammopathiesClinical Chemistry · 2026
- Anti-CD38 monoclonal antibodies in multiple myeloma and beyond: immunotherapeutic mechanisms, evidence maturity, and clinical positioning.Frontiers in immunology · 2026