B-044 Investigation of the Effect of Shortening Centrifugation Time on Hormone Test Results with the Transition to DxA 5000 Automation System

Oytun Portakal, Samil Binici, Meryem Zengin Bulbul, Bedri Bera Eken, Nazende Islak, Incilay Lay, Selçuk Dagdelen, Z Gunnur Dikmen

Observational · n=40

Key points

From the paper's conclusion
  • This study aimed to determine whether reducing centrifugation time from 6 to 4 minutes produces statistically and clinically significant differences in selected immunoassay parameters.

Abstract

Background Serum hormone measurements are highly dependent on preanalytical variables, particularly centrifugation time, speed, and temperature, which directly influence sample quality and analytical reliability. In February 2025, our laboratory transitioned from the Beckman Coulter Power Processor system to the DxA 5000 automation platform. With this upgrade, centrifugation time was reduced from 6 minutes to 4 minutes at 4000 rpm. However, the clinical impact of this modification has not been systematically evaluated. This study aimed to determine whether reducing centrifugation time from 6 to 4 minutes produces statistically and clinically significant differences in selected immunoassay parameters. Methods This prospective method-comparison study was conducted in the Central Laboratory of Hacettepe University Hospitals. Venous blood samples were simultaneously collected into two serum separator tubes (SST) from 40 volunteers. Samples were assigned to: Group A: tubes centrifuged at 4000rpm for 6 minutes and Group B: tubes centrifuged at 4000 rpm for 4 minutes. All other preanalytical and analytical conditions (clotting time, analyzer, processing sequence, testing time, and temperature) were standardized. Samples were analyzed on the same DxI800 immunoassay analyzer for thyroid function tests (fT4, TSH) and anemia-related parameters (ferritin, folate, vitamin B12). Statistical comparison between groups was performed using the Mann–Whitney U test. Passing–Bablok regression analysis was applied to evaluate systematic and proportional bias. Agreement between methods was assessed using Bland–Altman analysis. Concordance was quantified using Lin’s Concordance Correlation Coefficient (CCC). Results No statistically significant differences were observed between the two centrifugation protocols for TSH (p = 0.84), fT4 (p = 0.78), vitamin B12 (p = 0.83), folate (p = 0.65), or ferritin (p = 0.69). Passing–Bablok regression demonstrated slopes not significantly different from 1 and intercepts not significantly different from 0 for all analytes, indicating absence of systematic or proportional bias. CCC values exceeded 0.95 for all parameters, indicating excellent concordance. Bland–Altman analysis showed that mean bias was clinically negligible and total error remained below allowable total error (TEa) limits for all analytes Conclusion Shortening centrifugation time from 6 minutes to 4 minutes at 4000rpm does not produce statistically or clinically significant differences in the evaluated immunoassay parameters. The two centrifugation protocols are analytically interchangeable. Implementation of the 4-minute protocol within the DxA 5000 automation system is therefore appropriate and may contribute to improved laboratory workflow efficiency without compromising result reliability.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers