B-031 A Next-generation Assay for Rheumatoid Factor with Enhanced Performance on the Maccura C1000 Automatic Biochemistry System
Yalin Mao, Fu Pang, Changying Wu, Pingping Gan, Xiaohui Zhan, Renlin Li
Key points
From the paper's conclusion- The present study aimed to systematically evaluate the analytical performance of this new-generation RF assay on the Maccura C1000 Automatic Biochemistry System, including precision, on-board stability, linear range, hook effect, and method comparison.
Abstract
Background Rheumatoid factor (RF) is a key serological marker for the diagnosis and prognostic evaluation of rheumatoid arthritis. Currently, conventional immunoturbidimetric assays generally exhibit poor precision and on-board stability, whereas standard latex-enhanced immunoturbidimetric assays are limited by a narrow analytical measurement range. The narrow range results in high retesting rates in clinical practice, significantly prolonging sample turnaround time and increasing operational costs. To address these challenges, we optimized the coupling efficiency of core latex particles and the overall reaction formulation to develop a new-generation RF assay reagent (RF Gen.2). The present study aimed to systematically evaluate the analytical performance of this new-generation RF assay on the Maccura C1000 Automatic Biochemistry System, including precision, on-board stability, linear range, hook effect, and method comparison. Methods All above performance evaluations of our RF Gen.2, based on the principle of latex-enhanced immunoturbidimetry, were conducted on the Maccura C1000 System in accordance with CLSI guidelines. Firstly, linearity performance was evaluated according to CLSI EP06-A. Then, repeatability and within-laboratory precision were assessed according to CLSI EP15-A3 using two levels of pooled human sera (25 IU/mL; 50 IU/mL). Next, for hook effect evaluation of the new-generation RF assay, high-concentration clinical samples were analyzed to determine the highest analyte concentration threshold at which the hook effect would not interfere with the results. And, the new-generation RF reagent was compared in parallel with four mainstream commercially available reagents (Manufacturers W and D: immunoturbidimetric assays; Manufacturers R and B: latex-enhanced immunoturbidimetric assays) on their respective matched systems. The comparison focused on linearity range, repeatability, within-laboratory precision, and hook effect threshold. Finally, the method comparison was evaluated according to CLSI EP09-A3 by testing 100 clinical samples, using the Manufacturer W reagent as the comparator method. Results Integrated with Maccura C1000 System, the new-generation RF assay demonstrated excellent analytical performance, as shown in Table 1. And the RF assay also demonstrated a strong correlation with the comparator method (y=-2.5757+1.0047x, r=0.9963, N=100). Conclusion The new-generation RF assay on the Maccura C1000 system exhibited a wider linearity range and superior resistance to the hook effect. In combination with its excellent precision and on-board stability, we conclude that this RF assay is highly suitable for meeting the high-throughput testing requirements of routine clinical laboratories.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- B-273 Study on the Application Effect of a Novel Serum Index Detection Method Based on Reagent BackgroundClinical Chemistry · 2026
- B-176 Evaluation of a Multiplex Rheumatoid Arthritis Serology Panel for Clinical Laboratory Integration: Concordance with Existing RF and Anti-CCP AssaysClinical Chemistry · 2026
- Multicenter evaluation of analytical performance and onboard stability of Mindray 10 concentrated biochemical reagentsJournal of Laboratory Medicine · 2026
- B-138 Implementing Likelihood Ratios in Autoimmune Diagnostics: Early Experiences from Laboratories and CliniciansClinical Chemistry · 2026
- CBC‑derived inflammatory indices for rheumatoid arthritis diagnosis and activity assessment: differential performance by serostatusFrontiers in Immunology · 2026
- B-214 Analytical Performance Evaluation of the ST2 Assay on the Maccura Immunoassay Platform in Dual-ModeClinical Chemistry · 2026