B-214 Analytical Performance Evaluation of the ST2 Assay on the Maccura Immunoassay Platform in Dual-Mode
Junxi Tian, Xue Xia, Rong Hu, Yu Xu, Changgeng Chen, Renlin Li
Key points
From the paper's conclusion- Conclusion Maccura’s ST2 assay demonstrates excellent analytical performance, characterized by high sensitivity, precision, specificity, and strong concordance with a reference method.
Abstract
Background Growth stimulation expressed gene 2 (ST2) is a member of the interleukin-1 receptor family, with two isoforms linked to cardiac disease progression: soluble ST2 (sST2) and membrane-bound ST2 (ST2L). It is a valuable biomarker for risk stratification and prognosis in patients with cardiac diseases. This research presents an evaluation of the analytical performance of the ST2 assay manufactured by Maccura Biotechnology Co., Ltd. (Chengdu, China). Methods To accommodate different clinical scenarios, the platform provides dual-mode for ST2 Assay, including a routine mode with a 15-minute turnaround time and a short turnaround time (STAT) mode that delivers results within 9 minutes. The analytical performance of both modes has been validated following relevant CLSI guidelines. Among them, analytical sensitivity (LoB, LoD, LoQ) was determined following CLSI EP17-A2. LoQ was defined as the lowest concentration meeting predefined imprecision criterion. In addition, repeatability and within-laboratory precision were assessed according to CLSI EP05-A3. Patient pools of ST2 were measured in duplicate, and two runs were performed per day for 20 days. Analytical specificity, including cross-reactivity and interference, was evaluated following CLSI EP07-A2. Method comparison was performed against an established assay on the Mindray system according to CLSI EP09-A3. A total of 86 clinical samples covering a wide concentration range were analyzed, which were classified as positive or negative based on the clinical decision threshold. Agreement was assessed using Passing-Bablok regression for quantitative comparison and Kappa statistics for qualitative agreement. Results The analytical sensitivity for both modes is shown in Table 1, with each demonstrating favorable performance. Regarding precision, the routine mode showed repeatability and within-laboratory precision of 1.67%-2.04% and 2.39%-3.14%, respectively, while the corresponding values for the STAT mode were 1.37%-1.93% and 2.12%-3.03%. For the specificity, the assay exhibited no cross-reactivity with recombinant IL-1R1, IL-1a, IL-1ß, cTnI, or NT-proBNP, in both modes. And, no significant interference was observed in either assay mode when 33 cardiovascular drugs were spiked into healthy human serum. Regarding method comparison, Passing-Bablok regression yielded the equation y = 2.507+0.936x (Pearson r = 0.992) for the routine mode and y = 2.583 + 0.957x (Pearson r = 0.992) for the STAT mode. Both modes demonstrated exceptionally high concordance with the established assay. In qualitative analysis, the agreement rates between the routine mode and the Mindray system were 98.5% for positive samples, 94.7% for negative samples, and 97.7% overall, and the corresponding values for the STAT mode were 100%, 94.4%, 98.8%, respectively. Conclusion Maccura’s ST2 assay demonstrates excellent analytical performance, characterized by high sensitivity, precision, specificity, and strong concordance with a reference method. Its STAT mode provides accurate results within 9 minutes, enabling rapid risk stratification for acute cardiovascular patients.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- The diagnostic and prognostic value of growth stimulation expressed gene 2 in patients with coronary artery diseaseEuropean Heart Journal · 2026
- A-006 Diagnostic performance of an investigational high-sensitivity cardiac troponin T and long-T ITC troponin complex assays for acute myocardial infarction in females and malesClinical Chemistry · 2026
- Dual-channel surface-enhanced Raman immunoassay for simultaneous determination of atrial-fibrillation-associated NT-proBNP and high-sensitivity cardiac troponin I in artificial serum.Analytical methods : advancing methods and applications · 2026
- The prognostic value of circulating soluble ST2 in patients with chronic heart failureJournal of International Medical Research · 2026
- B-030 Analytical Performance Evaluation of the Maccura Glycosylated Hemoglobin Analyzer G 02 (High-Performance Liquid Chromatography)Clinical Chemistry · 2026
- Stacking Ensemble Learning for Diagnosing Acute Myocardial Infarction in Mice Using Plasma Raman Spectroscopy and Mass Spectrometry.Journal of cardiovascular translational research · 2026