B-214 Analytical Performance Evaluation of the ST2 Assay on the Maccura Immunoassay Platform in Dual-Mode

Junxi Tian, Xue Xia, Rong Hu, Yu Xu, Changgeng Chen, Renlin Li

Key points

From the paper's conclusion
  • Conclusion Maccura’s ST2 assay demonstrates excellent analytical performance, characterized by high sensitivity, precision, specificity, and strong concordance with a reference method.

Abstract

Background Growth stimulation expressed gene 2 (ST2) is a member of the interleukin-1 receptor family, with two isoforms linked to cardiac disease progression: soluble ST2 (sST2) and membrane-bound ST2 (ST2L). It is a valuable biomarker for risk stratification and prognosis in patients with cardiac diseases. This research presents an evaluation of the analytical performance of the ST2 assay manufactured by Maccura Biotechnology Co., Ltd. (Chengdu, China). Methods To accommodate different clinical scenarios, the platform provides dual-mode for ST2 Assay, including a routine mode with a 15-minute turnaround time and a short turnaround time (STAT) mode that delivers results within 9 minutes. The analytical performance of both modes has been validated following relevant CLSI guidelines. Among them, analytical sensitivity (LoB, LoD, LoQ) was determined following CLSI EP17-A2. LoQ was defined as the lowest concentration meeting predefined imprecision criterion. In addition, repeatability and within-laboratory precision were assessed according to CLSI EP05-A3. Patient pools of ST2 were measured in duplicate, and two runs were performed per day for 20 days. Analytical specificity, including cross-reactivity and interference, was evaluated following CLSI EP07-A2. Method comparison was performed against an established assay on the Mindray system according to CLSI EP09-A3. A total of 86 clinical samples covering a wide concentration range were analyzed, which were classified as positive or negative based on the clinical decision threshold. Agreement was assessed using Passing-Bablok regression for quantitative comparison and Kappa statistics for qualitative agreement. Results The analytical sensitivity for both modes is shown in Table 1, with each demonstrating favorable performance. Regarding precision, the routine mode showed repeatability and within-laboratory precision of 1.67%-2.04% and 2.39%-3.14%, respectively, while the corresponding values for the STAT mode were 1.37%-1.93% and 2.12%-3.03%. For the specificity, the assay exhibited no cross-reactivity with recombinant IL-1R1, IL-1a, IL-1ß, cTnI, or NT-proBNP, in both modes. And, no significant interference was observed in either assay mode when 33 cardiovascular drugs were spiked into healthy human serum. Regarding method comparison, Passing-Bablok regression yielded the equation y = 2.507+0.936x (Pearson r = 0.992) for the routine mode and y = 2.583 + 0.957x (Pearson r = 0.992) for the STAT mode. Both modes demonstrated exceptionally high concordance with the established assay. In qualitative analysis, the agreement rates between the routine mode and the Mindray system were 98.5% for positive samples, 94.7% for negative samples, and 97.7% overall, and the corresponding values for the STAT mode were 100%, 94.4%, 98.8%, respectively. Conclusion Maccura’s ST2 assay demonstrates excellent analytical performance, characterized by high sensitivity, precision, specificity, and strong concordance with a reference method. Its STAT mode provides accurate results within 9 minutes, enabling rapid risk stratification for acute cardiovascular patients.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

B-214 Analytical Performance Evaluation of the ST2 Assay on the Maccura Immunoassay Platform in Dual-Mode — Science Skim