A-180 Measurement of hemoglobin A1c across three methodologies: A comparison in the presence and absence of hemoglobin variants
Marvin Berman, Casey Williamson, Pow Cheng, Xiao Wang, Davor Brinc, Vathany Kulasingam
Key points
From the paper's conclusion- This study compared the HbA1c measurements obtained in the presence and absence of hemoglobin variants using three platforms which have distinct analytical principles.
Abstract
Background Hemoglobin A1c (HbA1c) is a standardized biomarker used in the diagnosis and management of diabetes. Since HbA1c cutoffs are universally applied across clinical laboratories, agreement between analytical platforms is essential. Methodological differences among systems (chromatographic, enzymatic, and immunochemical) may lead to clinically meaningful variability. In addition, developmental and pathological hemoglobin variants, including elevated fetal hemoglobin (HbF), may differentially affect HbA1c methodologies. This study compared the HbA1c measurements obtained in the presence and absence of hemoglobin variants using three platforms which have distinct analytical principles. Methods HbA1c (%) was measured using the Bio-Rad D-100 (high-performance liquid chromatography; reference method for this study), Abbott Alinity c (enzymatic assay) and Roche cobas c503 (immunoassay). Cohort 1 included 50 individuals with HbA1c values (3.5–5.9%) and no hemoglobin variants. Cohort 2 included 140 individuals with hemoglobin variants (3.5–10.4%), characterized by the Bio Rad D 100 or Sebia Capillarys systems. Variants included hemoglobinopathies/thalassemias (22 cases) and/or elevated HbF (5–30%). Agreement between methods was assessed using Deming regression and Bland–Altman plots with D-100 as the comparator. For samples with isolated HbF elevation (excluding structural variants), linear regression evaluated the effect of HbF (%) on HbA1c bias for Alinity and Roche. The percentage of reported or out-of-range/suppressed values for hemoglobin variants was compared. Results In cohort 1, both Alinity and Roche showed strong agreement with D-100. Roche demonstrated average relative bias 1.11%; correlation coefficient r=0.938 while Alinity exhibited average relative bias 2.78%; r=0.938. In cohort 2, greater bias was observed with the presence of variants: Roche (bias -5.85%; r=0.954) and Alinity (bias -3.88%; r=0.960). For samples exhibiting only elevated HbF, both systems demonstrated a negative proportional effect relative to D-100 (Alinity: -0.052% HbA1c per 1% HbF; Roche: -0.057% per 1% HbF). Among variant samples with known hemoglobinopathies/thalassemias, D-100 suppressed results for 15.7% of cases (manually performed in lab) while Alinity did not report a specific value due to results being out of range in 12.9%, and Roche reported values for all samples. Conclusion The D-100, Alinity, and Roche platforms demonstrated high comparability in HbA1c samples without hemoglobin variants. In the presence of variants, particularly elevated HbF, both Alinity and Roche showed modest negative proportional bias relative to D-100 but maintained a strong correlation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- B-233 Haemoglobin Variants and Their Impact on HbA1c Measurement: Experience with Two Routine MethodsClinical Chemistry · 2026
- 70 Bio-Rad D-100 HbA1c Assay in Variant Populations: A Critical Analysis of Flagging and Reporting IssuesAmerican Journal of Clinical Pathology · 2026
- A-136 Improving HbA1c Result Reportability Using Dual-Method Verification: Bio-Rad D-100 HPLC and Siemens Healthineers DCA ImmunoassayClinical Chemistry · 2026
- B-011 Evaluation of the Tosoh HLC-723GR01 HbA1c Analyzer Using Normal Hemoglobin Samples and in the Presence of Common Hemoglobin and Thalassemic VariantsClinical Chemistry · 2026
- B-299 Analytical performance, variant detection and result reportability of the CAPILLARYS 3 TERA HbA1c assay relative to the Siemens Atellica CH930Clinical Chemistry · 2026
- Glycated Haemoglobin Versus Serum Fructosamine in Assessing Glycaemic Control Amongst Type 2 Diabetes Mellitus Patients With Haemoglobin Variants: A Systematic Review.BioMed research international · 2026