Creatinine Quantification Using a Low‐Cost Portable Spectrophotometric Device
Bini Abraham, Praveena Venugopal, Karuvath Yoosaf
Key points
From the paper's conclusion- The system exhibits an excellent linear relationship between the absorbance and creatinine concentration ( R 2 = 0.999; slope = 0.06745), matching that of the conventional spectrophotometric device.
Abstract
Chronic kidney disease (CKD) is a serious health concern that occurs when the kidneys cannot filter blood properly, and creatinine is one of the main biomarkers for kidney dysfunction. The Jaffe method is the most widely practiced as it is cost‐effective, simple, and adaptable for routine laboratory use. To facilitate CKD's point‐of‐care diagnosis, we developed a low‐cost device integrating a blue LED light source, a light‐dependent resistor (LDR) detector, and a microcontroller unit, and it operates based on the principles of the Jaffe assay by measuring optical density at 463 nm. The ESP32 microcontroller's built‐in wi‐fi module enabled seamless data transfer to a computer or mobile phone for further analysis. The system exhibits an excellent linear relationship between the absorbance and creatinine concentration ( R 2 = 0.999; slope = 0.06745), matching that of the conventional spectrophotometric device. Validation experiments utilizing samples in water, artificial urine, and artificial serum demonstrated low limits of detection (∼0.029 mg/dL) and quantification (∼0.098 mg/dL). The device demonstrated strong predictive performance with RMSE values of 0.206 mg/dL for water samples and ∼ 0.3 mg/dL for artificial urine and artificial serum samples. High correlation coefficients ( R 2 > 0.99) across matrices confirm the system's good accuracy, precision, and reliability.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Development of a Low-Cost Portable Exhaled Breath Ammonia Detector for Supplementary Five-Stage CKD Classification Using Embedded Threshold Logic.Sensors (Basel, Switzerland) · 2026
- Analytical Evaluation of Simulated Kidney-Disease Profiles: Comparison of a Portable Near-Infrared Spectrophotometer with a Benchtop FT-NIR.Biosensors · 2026
- Development and Analytical Validation of a Smartphone‐Based Quantitative Lateral Flow Immunoassay for Serum Cystatin‐CJournal of Clinical Laboratory Analysis · 2026
- Sample-to-Answer Quantitative Rapid Test for Point-of-Care Monitoring of Urinary Cystatin C Levels.ACS sensors · 2026
- Transdermal Creatinine Sensing in Live Animals with an Integrated Electrochemical Microneedle Patch.ACS sensors · 2026
- A Portable, Low-Cost Spectrophotometer for Absorbance and Selected-Wavelength Fluorescence Measurements.Journal of analytical methods in chemistry · 2026