Clinical Chemistry

A-353 Relationship between ionized calcium and pH among different sample types and analyzers

Ibrahim Hashim, Rebecca Kim, Kibibi Smith

Key points

From the paper's conclusion
  • The venous samples pH-ionized calcium relationship observed when using the ABL instrument was significantly different (p <0.05) from that observed when using the iSTAT® point of care instrument.

Abstract

Background Circulating concentration of ionized calcium is known to be influenced by pH and that hydrogen ions displace calcium from its binding proteins. It is not known if the relationship is influenced by sample type nor by the instrument used for their measurements. This study compared ionized calcium levels and its associated pH in arterial and venous samples obtained by laboratory-based and by point of care instruments. Methods Reported ionized calcium and associated pH results obtained using blood gases instrument, ABL90 Flex Plus,(Radiometer America Inc., Brea, CA, USA) and point of care device iSTAT® (Abbott Diagnostics, NJ, USA) were retrospectively obtained from the electronic medical records for the period, January 2026. iSTAT® uses two separate cartridges namely CHEM8+ and EG7+ for ionized calcium and pH respectively. Statistical analysis was performed using NCSS® 2025, statistical software (NCSS, Utas, USA) Results Ionized calcium and associated pH results reported in a total of 1208 blood gases samples (1145 arterial, 48 venous, and 15 mixed venous-arterial) were obtained during the study period. Among arterial samples pH and associated ionized calcium levels ranged from 6.93 to 7.2 (median 7.37), and from 2.26 to 7.35 (median 4.56 mmol/L) respectively when performed by ABL-instrument. The relationship between ionized calcium was found to be governed by the formula; iCa = (10.4750) + (-0.8016) * pH. Among venous samples, pH and associated ionized calcium levels ranged from 7.09 to 7.53 (median 7.37), and from 3.25 to 5.61 (median 4.61 mmol/L) respectively when performed by ABL-instrument. The relationship when using ABL venous samples was governed by the formula; iCa= (0.9461) + (0.4941) *pH Among venous samples, pH and associated ionized calcium levels ranged from 6.81 to 7.33 (median 7.02), and from 4.4 to 6.6 (median 4.9 mmol/L) respectively when performed by iSTAT®-instrument. The relationship between ionized calcium and pH when using iSTAT® venous samples was governed by the formula; iCa = (23.3021) + (-2.5576) * pH. The venous samples pH-ionized calcium relationship observed when using the ABL instrument was significantly different (p <0.05) from that observed when using the iSTAT® point of care instrument. This may be a reflection of the differences in measurement methodologies. Conclusion Results showed negative correlation and negative slope relationship between ionized calcium and pH indicating displacement of calcium from its binding proteins in the presence of high hydrogen ions concentration (low pH). Furthermore, venous blood ionized calcium and pH values exhibited a higher intercept and higher negative slope when performed by the point of care device compared with laboratory-based instrument.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers