A-198 Performance validation of an updated osmometer for serum/plasma and urine testing and comparison to an established osmometer

J Alan Erickson, Elizabeth Frank

Key points

From the paper's conclusion
  • Percent biases were 0.7 and 1.1% and r2 values of 0.9998 and 0.9997 for serum/plasma and urine, respectively. (Deming regression, Bland-Altman analysis, n = 46 each specimen type.) Conclusions: The updated osmometer demonstrates satisfactory performance for measuring osmolality in human serum/plasma and urine.

Abstract

Background Osmolality, the measurement of total solute concentration in an aqueous solution, is based on the concept that the solute concentration is proportional to the freezing point of the solution. Clinical laboratory testing of osmolality in human plasma, serum, and urine typically employs freezing-point depression. Osmolality is used to evaluate and monitor electrolyte and acid-base disorders, including water balance (dehydration, overhydration), kidney function, diabetes, poisoning (ethanol, methanol, ethylene glycol), and sodium disorders (hyponatremia, hypernatremia). Results are expressed as milliosmoles per kilogram of water (mOsm/kg H2O). The OsmoPRO MAX Automated Osmometer (Advanced Instruments) was introduced in 2022. In early 2024, the instrument was recalled voluntarily by the manufacturer because of recurrent system error messages and the need for frequent calibration. After implementing technical improvements and software updates, the OsmoPRO MAX was reintroduced in 2025. Our goal was to validate two OsmoPRO MAX instruments for implementation in our laboratory. Methods Specimens were collected according to the University of Utah’s Institutional Review Board approved protocols. Testing was conducted according to the instrument manufacturer’s protocols. Two osmometers were validated for both serum/plasma and urine specimen types. The performance characteristics evaluated included linearity, analytical sensitivity, imprecision, analyte recovery, carryover, and method comparison against the established A2O Advanced Automated Osmometer (A2O). Results Linearity studies produced a slope of 1.003, y-intercept of 0.618 and r2 of 1.000 for osmometer 1. For osmometer 2, the slope, y-intercept and r2 were 0.9965, 1.262 and 1.000, respectively. (Osmolality linearity kit, n = 5, linear regression, 0 – 2000 mOsm/kg H2O.) The analytical sensitivity was 0 mOsm/kg H2O for both osmometers. For osmometer 1, repeatability CVs were 0.2 – 0.3% and within laboratory CVs 0.2 – 1.9%. For osmometer 2, CVs were 0.1 – 0.4% and 0.2 – 2.1% for repeatability and within laboratory CVs, respectively. (Six concentrations, four replicates per day per concentration level over 5 days.) Analyte recoveries ranged from 100.2 – 100.5% and 99.7 – 100.1% for osmometers 1 and 2, respectively. Carryover studies resulted in 0.25 mOsm/kg H2O for osmometer 1; no carryover (0.00 mOsm/kg H2O) was observed for osmometer 2. For osmometer 1, a method comparison study against the A2O osmometer generated slopes of 1.013 and 1.016 for serum/plasma and urine, respectively. The intercepts were -1.755 and -1.249, respectively. Percent biases were 0.7% and 1.2%, for serum/plasma and urine, respectively, with r2 = 0.9998 for both specimen types. For osmometer 2 vs. the A2O, the slopes were 1.011 and 1.009 for serum/plasma and urine, with intercepts of -1.258 and 1.584, respectively. Percent biases were 0.7 and 1.1% and r2 values of 0.9998 and 0.9997 for serum/plasma and urine, respectively. (Deming regression, Bland-Altman analysis, n = 46 each specimen type.) Conclusions: The updated osmometer demonstrates satisfactory performance for measuring osmolality in human serum/plasma and urine. The issues leading to the voluntary recall of the OsmoPRO MAX Automated Osmometer have been successfully resolved. The OsmoPRO MAX is acceptable for assessing patient conditions and disorders affecting osmolality.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

A-198 Performance validation of an updated osmometer for serum/plasma and urine testing and comparison to an established osmometer — Science Skim