Clinical Chemistry

A-188 Analytical Validation of Albumin Measured in Pericardial Fluid

Abbie Anderson, Leslie Donato, Vlad Vasile, Darci Block, Jeffrey Meeusen

Key points

From the paper's conclusion
  • Conclusion These data show that albumin can be precisely measured in pericardial fluid within a linear range of 0.5-5.6 g/dL.

Abstract

Background Albumin measurement in pericardial fluid is frequently requested in our reference laboratory practice. The serum to ascites albumin gradient (SAAG) is used to distinguish between transudative and exudative effusions. However, growing evidence suggests pericardial albumin is less useful in this scenario. Our objective was to validate analytical accuracy of albumin measurement in pericardial fluid and compare frequency of transudate classifications. Methods Albumin concentrations were measured using a Roche Cobas c501, (Roche, Indianapolis, IN). Within-laboratory coefficient of variation (CVWL) and repeatability (CVr) were determined by repeated measurement of pooled clinical residual pericardial fluid. Linearity was assessed by mixing pericardial fluid pools with high and low albumin concentrations. Dilutional recovery from clinically ordered pericardial fluid albumin analyses (n=772) were used to determine analytical accuracy. SAAG transudate diagnostic threshold of 1.1 g/dL (peritoneal) or 1.2 g/dL (pleural) was evaluated in clinical data from patients with paired serum and body fluid albumin measured within a 2-day interval (n=3,031). Results Albumin imprecision in pericardial fluid was assessed at 1.4 g/dL (CVWL = 6.3%, CVr = 5.7%) and 3.1 g/dL (CVWL = 2.2%, CVr = 2.2%). Mixing studies confirmed linearity between 0.5-5.6 g/dL (slope 1.02, R2 = 0.999). Dilutional recovery was performed in 772 clinical orders spanning 0.9 g/dL to 4.8 g/dL, the Passing-Bablok fit of expected versus measured values had slope = 1.09 g/dL and intercept = -0.2 g/dL. In the clinical data, serum albumin was similar for patients in all cases. Pericardial albumin [median(IQR) 2.2(1.0-2.6) g/dL] was significantly higher compared to peritoneal [1.0(0.5-1.7) g/dL; p<0.001] or pleural fluid [1.6(1.0-2.2) g/dL; p<0.001], leading to significantly lower SAAG calculation for pericardial [1.3(0.5-1.9) g/dL] versus peritoneal [1.9(1.4 – 2.4); p<0.001] or pleural fluid [1.5(1.0-2.1); p<0.001]. Applying the SAAG threshold of 1.2 g/dL classified 84% of peritoneal fluids and 58% of pericardial fluids as transudative effusions. The Light’s criteria threshold of 1.2 g/dL led to 65% of pleural fluids and 55% of pericardial fluids being classified as transudative. Conclusion These data show that albumin can be precisely measured in pericardial fluid within a linear range of 0.5-5.6 g/dL. In clinical cases of evaluating SAAG ratio in body fluid samples, albumin concentrations are generally higher in pericardial fluid compared to peritoneal or pleural fluids. This increased pericardial albumin concentration leads to lower SAAG calculations. Application of typical transudate versus exudate thresholds in pericardial fluids leads to more exudative determinations. This data suggests that pericardial fluid may warrant a unique diagnostic threshold to be determined in further studies.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers