A-179 Development and Evaluation of a Novel Candidate Reference Method for the Simultaneous Determination of Serum Free Thyroxine (FT4) and Free Triiodothyronine (FT3) Using Equilibrium Dialysis Combined with Isotope Dilution LC-MS/MS

Weiyan Zhou

Key points

From the paper's conclusion
  • Background To develop and validate a novel candidate reference measurement procedure (cRMP) for free thyroxine (FT4) and free triiodothyronine (FT3) using volume-adjustable equilibrium dialysis (ED) plates, thereby providing a practical and highly sensitive solution for the global standardization of FT4 and FT3 measurements.

Abstract

Background To develop and validate a novel candidate reference measurement procedure (cRMP) for free thyroxine (FT4) and free triiodothyronine (FT3) using volume-adjustable equilibrium dialysis (ED) plates, thereby providing a practical and highly sensitive solution for the global standardization of FT4 and FT3 measurements. Methods We established an isotope dilution liquid chromatography-tandem mass spectrometry (LC-MS/MS) method for the simultaneous determination of FT4 and FT3. The ED conditions were meticulously optimized using adjustable ED plates. The method was strictly validated in accordance with international guidelines. Results The optimized ED conditions were as follows: the ratio of serum to dialysis buffer could be adjusted according to the sample volume or the sensitivity of the mass spectrometer. On the ABSciex 6500+ instrument, 500 µL of serum to 1500 µL of buffer at 37°C for 18 hours. The within-day, between-day, and total imprecision (CV) of FT4 and FT3 were =3.0%, =3.6%, and =4.0%, respectively. The average recovery of spiked samples in the dialysis buffer was between 99.03% and 104.26%. The relative matrix effect after internal standard correction was 1.63% - 1.66%. Accuracy was verified using serum materials assigned values by Ghent University, with deviations less than 2.61%. Conclusion This study presents a robust, fully validated, and reproducible cRMP for FT4 and FT3. The method offers the advantages of adjustable sample-to-dialysis buffer ratio, simplified operation, significantly improved sensitivity, and reduced sample volume, providing a feasible approach for the standardization of FT4 and FT3.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers