A candidate reference measurement procedure based on isotope dilution-liquid chromatography-tandem mass spectrometry for standardizing human serum D-α-tocopherol measurements.

Wei Jiang, Bingling Gao, Tao Yang, Liping Tang, Keqi Sun, Peng Xiao

Reference System Center, Maccura Biotechnology Co., Ltd., Chengdu 610219, China. · Reference System Center, Maccura Biotechnology Co., Ltd., Chengdu 610219, China. Electronic address: sunkeqi@maccura.com.

Key points

From the paper's conclusion
  • A reliable ID-LC-MS/MS-based cRMP was developed, which provided high specificity, accuracy and precision for D-α-tocopherol quantification, contributing to the D-α-tocopherol measurement standardization and traceability.

Abstract

BACKGROUND: Vitamin E deficiency commonly occurs in infants and often manifests clinically as anemia, edema, nervous system abnormalities, and other symptoms. D-α-tocopherol, the predominant and most biologically active form of vitamin E in human serum. In this study, we developed and validated a candidate reference measurement procedure (cRMP) for human serum D-α-tocopherol using isotope dilution liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS). METHODS: Isotopically labeled internal was dried under nitrogen at 40 °C before adding appropriate amount of serum sample. The mixture was processed by protein precipitation with methanol and liquid-liquid extraction with n-hexane. The extraction was dried, reconstituted, and analyzed by LC-MS/MS. The cRMP was evaluated according to the Clinical and Laboratory Standards Institute (CLSI) approved guidelines. RESULTS: For samples ranging from 17.0 to 43.6 μmol/L, intra-batch imprecision were below 2.6% and inter-batch imprecision were below 2.0%. The limit of detection and lower limit of measuring interval were determined as 0.0119 and 0.662 μmol/L respectively. Analysis of NIST Standard Reference Material (SRM) 968f resulted in biases of 0.8% for Level 1and 0.5% for Level 2. Recovery rate ranged from 98.1% to 102.5%. Excellent linearity was observed across a range from 0.440 to 78.7 μmol/L. Linear dependency was observed between three clinical routine methods and this cRMP. The method demonstrated good agreement in both method comparison and interlaboratory comparison studies. CONCLUSIONS: A reliable ID-LC-MS/MS-based cRMP was developed, which provided high specificity, accuracy and precision for D-α-tocopherol quantification, contributing to the D-α-tocopherol measurement standardization and traceability.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers