A-138 False-Positive Serum Qualitative hCG Results Associated with a Specific Lot of Greiner Bio-One VACUETTETM Red Gel Tubes Using the OSOM Ultra hCG Kit

Uttam Garg, Trang Nguyen, Amy Wiebold, Angela Ferguson

Key points

From the paper's conclusion
  • Conclusions A lot-specific pre-analytical interference associated with Greiner red gel blood collection tubes resulted in false-positive qualitative hCG results on the OSOM Ultra hCG assay.

Abstract

Background Qualitative human chorionic gonadotropin (hCG) assays are widely used for rapid pregnancy screening in clinical practice, particularly in emergency, surgery and other outpatient and inpatient settings. These assays are frequently utilized because they provide rapid results and are readily available when quantitative hCG testing is not immediately available. Both urine and serum specimens may be used for qualitative hCG testing. Although urine is generally the specimen of choice, serum is commonly used when urine samples cannot be obtained, such as in pediatric patients or critically ill individuals. False-positive qualitative hCG results are uncommon; however, they may occur due to pre-analytical interferences. Such false-positive results can lead to inappropriate clinical decisions, unnecessary diagnostic procedures, and patient anxiety, underscoring the importance of recognizing and investigating potential sources of assay interference. Objective To investigate a cluster of unexpected false-positive qualitative serum hCG results associated with samples collected in Greiner red gel blood collection tubes. Materials and Methods Blood was collected in Greiner Bio-One VACUETTETM red gel tubes or red no-gel tubes. Serum was separated from these tubes and tested for qualitative hCG using OSOM Ultra hCG Kit (SEKISUI Diagnostics) and quantitative VITROS total ß-hCG (Quidel Ortho Diagnostics). Results Three serum samples collected in Greiner red gel blood collection tubes (lot number B25063KD) from different patients produced unexpected positive results on the OSOM Ultra hCG qualitative assay. These samples were subsequently tested using the VITROS Total ß-hCG quantitative assay, and all three yielded negative results (<3 mIU/mL). To further investigate this discrepancy, an additional 26 randomly selected serum samples collected in the same Greiner red gel tube lot (B25063KD) were evaluated. Of these 26 samples, three produced positive results on the qualitative OSOM Ultra hCG assay; however, all corresponding quantitative measurements using the VITROS assay were negative (<3 mIU/mL). In addition, 53 serum samples collected using several other lots of Greiner gel tubes showed no false-positive qualitative hCG results. Furthermore, serum samples collected in no-gel tubes did not produce any false-positive results. Conclusions A lot-specific pre-analytical interference associated with Greiner red gel blood collection tubes resulted in false-positive qualitative hCG results on the OSOM Ultra hCG assay. Because false-positive results were observed in only a subset of samples from the affected tube lot, additional patient-specific factors, such as heterophilic antibody interference, may have contributed to the observed assay discordance.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers