A-109 Validation of a Multiplexed External Control Panel to Monitor the Performance of the Roche cobas® eplex respiratory pathogen panel 3

Nithya Sivaji, Jack Aldana Proulx, Priscilla Carnaroli, Joan Gordon, Matthew Held, Molly Kephart, Loren Krott, Daniel Magoon,

Computational · n=540

Key points

From the paper's conclusion
  • This external control provides a stable, easy-to-use solution for standardized training, validation, verification, environmental contamination detection, and routine monitoring of the performance of the Roche cobas® eplex RP3 panel.

Abstract

Background Acute respiratory infections pose a significant clinical challenge and place a substantial burden on the healthcare system. Patient diagnosis is often complicated due to similar symptoms caused by many viral and bacterial pathogens. Rapid and accurate identification is essential to improve diagnosis turnaround time, guide treatment decisions, and improve patient outcomes. Syndromic qualitative nucleic acid tests, including the cobas® eplex respiratory pathogen panel 3 (RP3) supports timely detection of a broad range of common respiratory illness organisms; however, ongoing assay performance monitoring is required to identify shifts, trends, and random errors due to variability in the test system. A non-infectious robust multiplexed external control was developed and validated to reliably monitor the detection of 16 viral and 4 bacterial pathogens assayed in the cobas eplex RP3 panel. Methods Non-infectious, multi-target constructs containing genome segments of all respiratory pathogens detected by the cobas eplex RP3 panel were designed in silico and ligated into engineered vectors to create stable frozen clones. In vitro DNA plasmids and RNA transcripts were generated, quantified and formulated in a proprietary matrix to stabilize and carry the nucleic acids throughout the entire testing process. Precision studies were performed to assess the reproducibility across multiple reagents and operators. Performance studies, including failure-mode, matrix-effect, shipping, and stability studies were conducted to confirm reliability and robustness. Results A total of 540 samples were tested across multiple reagent lots, instruments and operators. Of the 540 total runs tested, 520 were valid (96.3%) and used for analysis. For positive controls, 291/291 runs (100% true positive rate) resulted in correct detection calls for all multiplexed targets, and 228/229 negative control runs showed no false target detection (99.6% true negative rate). Thus, the overall correct call frequency of all runs was 99.8%. Failure mode studies demonstrated that the control effectively alerts end-users to system and operator errors and reliably detects analytical issues within the test system. Shipping studies confirmed robustness under simulated shipping delays and exposure to elevated temperatures, with ongoing real-time stability data presently supporting up to 15 months of refrigerated storage (2-8°C). Conclusion Validation studies demonstrated the reproducibility, precision, accuracy and robustness of the eplex RP3 Control M530. This external control provides a stable, easy-to-use solution for standardized training, validation, verification, environmental contamination detection, and routine monitoring of the performance of the Roche cobas® eplex RP3 panel.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers