Clinical Performance of a Fully Integrated Sample-to-Answer Multiplex PCR Respiratory Panel: A Multicenter Evaluation Using Prospective, Archived and Contrived Nasopharyngeal Specimens.
Chi-Sheng Tai, Hsing-Yi Chung, Tai-Han Lin, Chih-Kai Chang, Cherng-Lih Perng, Yifan Lin, Uanting Tsai, Matthew Beltrami,
Graduate Institute of Medical Science, National Defense Medical University, Taipei City 11490, Taiwan. · Division of Clinical Pathology, Department of Pathology, Tri-Service General Hospital, National Defense Medical University, Taipei City 11490, Taiwan.
Key points
From the paper's conclusion- For the remaining ten targets, prospective evidence was insufficient to confirm the acceptance criteria; archived and contrived results provide supportive evidence of detection but do not change this prospective classification.
Abstract
Background: Most syndromic respiratory PCR panels process one specimen per cartridge and instrument module. Objectives: We evaluated the agreement of a fully integrated, batch-capable sample-to-answer 22-target respiratory panel with BioFire FilmArray Respiratory Panel 2.1. Methods: Nasopharyngeal swabs were collected prospectively from symptomatic individuals aged ≥ 2 years at three US clinical sites between December 2025 and May 2026. Investigational and comparator testing were performed in separate, mutually blinded CLIA-certified laboratories; discordant specimens were retested with a third molecular assay, reported descriptively. Archived (n = 240) and contrived (n = 200) specimens for low-prevalence targets were analyzed separately. Positive and negative percent agreement (PPA, NPA) were reported with Wilson score 95% confidence intervals (CI). Results: Of 1501 participants enrolled, 1482 specimens yielded paired valid results; final invalid rates after repeat testing were 0.07% and 0.13%, respectively. Twelve targets met all prespecified prospective acceptance criteria (PPA 91.5-100%). Seven targets had PPA point estimates of 90.0-100% but lower 95% CI bounds below 80%, and three had no prospective positives. NPA was 99.6-100% for all 22 targets. Conclusions: The system showed high agreement with the comparator for targets well represented in prospective specimens. For the remaining ten targets, prospective evidence was insufficient to confirm the acceptance criteria; archived and contrived results provide supportive evidence of detection but do not change this prospective classification.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparative evaluation of Allplex RV master assay and biofire respiratory panel 2.1 plus for detection of respiratory virusesScientific Reports · 2026
- Development and Initial Analytical Evaluation of a Multiplex PCR-Based Targeted NGS Assay for Expanded Viral Detection in Clinical Respiratory Specimens.Diagnostics (Basel, Switzerland) · 2026
- Multicenter evaluation of Coyote FlashDetect LyocartE multiplex assays for six respiratory pathogens: a comparative analysis with Sansure and Cepheid assays.Microbiology spectrum · 2026
- Development of a multiplex real-time PCR assay with fluorescence probe-melting-curve analysis for one-tube detection of respiratory pathogens.PloS one · 2026
- Clinical and Analytical Validation of a Novel Switching RT-qPCR Assay for the Simultaneous Detection of Seven Respiratory Viruses.Journal of virological methods · 2026
- Syndromic diagnosis of upper respiratory infections in patients presenting with a suspicion of COVID-19 in Madagascar: contribution of the Biofire Film Array respiratory panel 2.1.BMC infectious diseases · 2026