Analytical Performance and Clinical Evaluation of a Fully Automated Multiplex RT-qPCR Platform for Rapid Detection of Central Nervous System Pathogens.
Mesut Yilmaz, Naim Mahroum
Department of Infectious Diseases and Clinical Microbiology, International School of Medicine, Istanbul Medipol University, 34810 Beykoz, Istanbul, Türkiye.
Key points
From the paper's conclusion- This study evaluated both the wet-lab analytical performance and the clinical performance evaluation of the Bioeksen Meningitis/Encephalitis Panel (BS-MEP) integrated onto the fully automated, high-throughput Sigmoida Lab platform.
Abstract
Meningitis and encephalitis necessitate rapid pathogen identification to guide therapy, as conventional methods are time-consuming. This study evaluated both the wet-lab analytical performance and the clinical performance evaluation of the Bioeksen Meningitis/Encephalitis Panel (BS-MEP) integrated onto the fully automated, high-throughput Sigmoida Lab platform. Analytical limits of detection (LoD) were defined via Probit analysis (95% threshold) by spiking negative cerebrospinal fluid (CSF) matrices. Target detection was verified using characterized reference materials for all 14 analytes, and in silico primer/probe coverage was assessed against taxon-specific sequence databases. Cross-reactivity was evaluated using high-prevalence non-target organisms. Clinical performance was evaluated retrospectively using 500 archived, anonymized CSF specimens from a single-center repository, with analyte-specific classifications compared with prespecified routine comparator methods. The automated platform provided an 80-min sample-to-result turnaround for up to 23 samples. Verified LoDs ranged from 472 to 2118 genome copies/mL. All inclusivity strains were successfully detected in 5/5 replicates. In silico coverage exceeded 98% combined, and zero wet-lab cross-reactivity was observed. Precision coefficients of variation (CVs) were consistently ≤1.38%. In the clinical evaluation, the pooled clinical agreement was high, with PPA ranging from 90.9% to 100% and NPA of 100% across the evaluated analytes. Analyte-level clinical sensitivity ranged from 90.9% to 100%, with zero false-positive results across all targets. The fully automated molecular system demonstrates excellent analytical robustness and strong clinical agreement for syndromic detection of major CNS pathogens. The automated workflow combines broad pathogen coverage with an approximately 80-min sample-to-result turnaround and warrants further prospective evaluation in routine clinical settings.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Multicenter evaluation of the multiplex polymerase chain reaction QIAstat-Dx Meningitis/Encephalitis Panel for the detection of central nervous system pathogens directly in the cerebrospinal fluid.Journal of clinical microbiology · 2026
- Quantitative and molecular adjudication of BioFire FilmArray and QIAstat-Dx discordance in meningitis/encephalitis diagnostics.Microbiology spectrum · 2026
- A-092 Application of the FilmArray® Meningitis/Encephalitis Panel in the Etiological Diagnosis of Suspected Meningitis: A Multiregional Study in BrazilClinical Chemistry · 2026
- FilmArray Meningitis/Encephalitis Panel accelerates central nervous system infection diagnosis and reduces empirical acyclovir use: A Japanese prospective study.Journal of infection and public health · 2026
- High-throughput screening for central nervous system infection-associated pathogens using TaqMan array cards, metagenomic next-generation sequencing, and targeted next-generation sequencing: a retrospective study.Frontiers in microbiology · 2026
- Evaluating a microfluidic loop-mediated isothermal amplification platform for rapid pathogen detection in endophthalmitis: a prospective studyJournal of Ophthalmic Inflammation and Infection · 2026