A-044 Reliable NT-proBNP Measurements from Capillary Blood: Enabling Decentralized Monitoring

Carlos Torres, Stacy Beal, Michael Rummel, Anissa Naranjo

Key points

From the paper's conclusion
  • Temperature chamber simulations confirmed the 3-day stability of capillary whole blood samples in microtainer tubes during fluctuating summer (10 to 42 ?) and winter (-4.0 to 35.6 ?) shipment extremes.

Abstract

Background Although traditionally used to rule out a heart failure diagnosis,the American Diabetes Association's recent guidelines now recommend consideration of screening diabetic patients with NT-proBNP measurement to aid in prevention of heart failure (“Cardiovascular Disease and Risk Management: Standards of Care in Diabetes,” 2026). NT-proBNP may also be followed in patients with established chronic kidney disease to help identify early heart failure, assess cardiovascular risk, and clarify the extent to which cardiac dysfunction is contributing to the patient’s clinical picture, such as fluid overload. Given the longer half-life of NT-proBNP compared with BNP and the need for repeat testing, use of decentralized specimen collection could be useful. As a first step towards the possibility of home collection, we aimed to validate the use of capillary blood collected either via fingerstick or the ImPress, a new upper arm capillary blood collection device. Methods The NT-proBNP assay was validated according to manufacturer’s instructions on the Roche e801 instrument using the Elecsys proBNP II test kit. In addition, matched venous, fingerstick, and ImPress samples were collected under IRB protocol from 10 healthy volunteers and 14 patients presenting to the emergency room for various conditions. Finally, stability studies were carried out. Results Sample site correlation: Venous vs. fingerstick (n=24) results (Range: 46.3 to 33156 pg/mL) yielded a slope of 1.026, y-intercept of -47.72, bias of 36.67 pg/mL (1.12%), and R-value of 0.9999. Venous vs. ImPress (n=24) results (Range: 46.3 to 33156 pg/mL) yielded a slope of 1.009, y-intercept of -9.71, bias of 20.36 pg/mL (0.62%), and R-value of 0.9997. Stability: Un-spun capillary whole blood was stable for 5 days at room temperature and 10 days refrigerated. Temperature chamber simulations confirmed the 3-day stability of capillary whole blood samples in microtainer tubes during fluctuating summer (10 to 42 ?) and winter (-4.0 to 35.6 ?) shipment extremes. Summer simulations (N=19) comparing pre- and post-shipment results (Range: 42.4 to 773 pg/mL) yielded a slope of 0.993, y-intercept of -9.9, bias of -11.24 pg/mL (-6.00%), and R-value of 0.9997. Winter simulations (N=20) (Range: 544 to 3777 pg/mL) showed a slope of 0.887, y-intercept of 19.4 pg/mL, bias of -187.1 pg/mL (-10.3%), and R-value of 0.9985. Conclusion Fingerstick and ImPress samples demonstrated excellent correlation with venous blood for NT-proBNP measurement. This validation has the potential to enable reliable, remote monitoring. Further studies with broader analytical ranges for NT-proBNP are currently being compiled to strengthen the rationale for at-home collection. Reference: American Diabetes Association Professional Practice Committee for Diabetes*, 10. Cardiovascular Disease and Risk Management: Standards of Care in Diabetes–2026. Diabetes Care 1 January 2026; 49 (Supplement_1): S216–S245. https://doi.org/10.2337/dc26-S010

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers