Alzheimer's Disease Biomarkers Measured by Capillary Blood Collection Versus Venous Plasma and Cerebrospinal Fluid.
Sara Ho, Ziling Shen, Alexandria Lewis, Megha Patel, Anny Zheng, Gianna Proscino, Thaovy Ho, Jiangxia Wang,
Johns Hopkins School of Medicine, Baltimore, Maryland, USA. · Johns Hopkins Bloomberg School of Public Health, Baltimore, Maryland, USA.
Key points
From the paper's conclusion- Neurodegeneration biomarkers Aβ42/40, NfL, and GFAP in capillary plasma (Tasso+, TAPII) strongly correlated with venous plasma measures. Capillary and venous pTau217 correlated weakly, indicating further assay refinement is needed. Tasso+ collection was less painful and caused less hemolysis compared to TAPII.
Abstract
OBJECTIVE: Blood biomarkers are increasingly critical for non-invasive Alzheimer's Disease (AD) diagnostics and research but typically require on-site processing. Remote point-of-care collection methods could enhance biomarker testing in clinical and research settings. We evaluated the safety and accuracy of blood collection via venous blood and two capillary blood collection devices (Tasso+ and TAPII) for measurements of amyloid-beta 42/40 ratio (Aβ42/40), phosphorylated-tau-217 (pTau217), neurofilament-light (NfL), and glial fibrillary acidic protein (GFAP) from capillary plasma to simultaneously collected venous plasma and cerebrospinal fluid (CSF). METHODS: Patients seen at the Johns Hopkins Center for CSF Disorders (N = 192) underwent diagnostic lumbar punctures and venipuncture. Warming packs were held to the skin for 5 min before collecting capillary blood (via TAPII or Tasso+ device) and numeric pain scores. CSF, venous, and capillary plasma were analyzed using Quanterix SIMOA immunoassays. Statistical analysis included Pearson/Spearman correlations, Lasso regression, and quantitative comparisons using Bland-Altman plots. RESULTS: 86 TAPII and 126 Tasso+ plasma samples were collected (77 M, 115 F; mean age 61 ± 18.3 years, range 16-91). Tasso+ devices were 90.3% successful in collecting > 50 μL whole blood, and TAPII devices were 86% successful. Mean pain scores were significantly higher in venous than capillary collections (Venous 2.24/10, TAPII 0.73/10, Tasso+ 0.2/10, p < 0.001). Capillary Aβ42/40, NfL, and GFAP strongly correlated with venous plasma analytes (Tasso+: Aβ42/40 r = 0.75, NfL r = 0.91, GFAP r = 0.80, all p < 0.001; TAPII: Aβ42/40 r = 0.74, NfL r = 0.86, GFAP r = 0.84, all p < 0.001). Tasso+ capillary pTau217 weakly correlated with venous (r = 0.23, p = 0.04). Capillary NfL and GFAP correlated modestly with CSF NfL and GFAP (TAPII NfL r = 0.74, GFAP r = 0.6, all p < 0.0001; Tasso+ NfL r = 0.69, GFAP r = 0.59, all p < 0.0001). Capillary Aβ42/40 and pTau217 did not significantly correlate with CSF measures. Lasso regression identified age and GFAP as significant predictors of AD across capillary and venous samples (Tasso+ AUC = 0.87, TAPII AUC = 0.75, Venous AUC = 0.83). INTERPRETATION: Neurodegeneration biomarkers Aβ42/40, NfL, and GFAP in capillary plasma (Tasso+, TAPII) strongly correlated with venous plasma measures. Capillary and venous pTau217 correlated weakly, indicating further assay refinement is needed. Tasso+ collection was less painful and caused less hemolysis compared to TAPII.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Performance of upper-arm capillary blood collection for Alzheimer's disease and central nervous system biomarkers: comparison of Tasso+ and venous plasma.medRxiv : the preprint server for health sciences · 2026
- Alzheimer’s Disease blood biomarkers measured through remote capillary sampling correlate with cognition in older adultsNature Communications · 2026
- Capillary brain-derived p-tau217 for remote triaging of Alzheimer's disease pathology: Clinical performance in a CLIA/CAP-certified laboratory.Alzheimer's & dementia : the journal of the Alzheimer's Association · 2026
- Analytical performance of a delayed-centrifugation blood collection tube for remote blood biomarker assessment.medRxiv : the preprint server for health sciences · 2026
- B-018 Diagnostic Performance and Clinical Utility of the Sysmex HISCL Automated Plasma Amyloid-ß 1-42/1-40 Assay in a Real-world SettingClinical Chemistry · 2026
- Delayed-centrifugation blood collection tube for remote blood biomarker assessment.medRxiv : the preprint server for health sciences · 2026